A fluid dynamics study in a 50 cc pulsatile ventricular assist device: Influence of heart rate variability

Jason C. Nanna, Michael A. Navitsky, Stephen R. Topper, Steven Deutsch, Keefe B. Manning

Research output: Contribution to journalArticle

10 Citations (Scopus)

Abstract

Although left ventricular assist devices (LVADs) have had success in supporting severe heart failure patients, thrombus formation within these devices still limits their long term use. Research has shown that thrombosis in the Penn State pulsatile LVAD, on a polyurethane blood sac, is largely a function of the underlying fluid mechanics and may be correlated to wall shear rates below 500 s -1. Given the large range of heart rate and systolic durations employed, in vivo it is useful to study the fluid mechanics of pulsatile LVADs under these conditions. Particle image velocimetry (PIV) was used to capture planar flow in the pump body of a Penn State 50 cubic centimeters (cc) LVAD for heart rates of 75-150 bpm and respective systolic durations of 38-50. Shear rates were calculated along the lower device wall with attention given to the uncertainty of the shear rate measurement as a function of pixel magnification. Spatial and temporal shear rate changes associated with data collection frequency were also investigated. The accuracy of the shear rate calculation improved by approximately 40 as the resolution increased from 35 to 12 m/pixel. In addition, data collection in 10 ms, rather than 50 ms, intervals was found to be preferable. Increasing heart rate and systolic duration showed little change in wall shear rate patterns, with wall shear rate magnitude scaling by approximately the kinematic viscosity divided by the square of the average inlet velocity, which is essentially half the friction coefficient. Changes in in vivo operating conditions strongly influence wall shear rates within our device, and likely play a significant role in thrombus deposition. Refinement of PIV techniques at higher magnifications can be useful in moving towards better prediction of thrombosis in LVADs.

Original languageEnglish (US)
Article number101002
JournalJournal of Biomechanical Engineering
Volume133
Issue number10
DOIs
StatePublished - Nov 10 2011

Fingerprint

Heart-Assist Devices
Hydrodynamics
Fluid dynamics
Shear deformation
Left ventricular assist devices
Heart Rate
Thrombosis
Rheology
Mechanics
Equipment and Supplies
Fluid mechanics
Polyurethanes
Velocity measurement
Friction
Biomechanical Phenomena
Viscosity
Pixels
Uncertainty
Heart Failure
Blood

All Science Journal Classification (ASJC) codes

  • Biomedical Engineering
  • Physiology (medical)

Cite this

Nanna, Jason C. ; Navitsky, Michael A. ; Topper, Stephen R. ; Deutsch, Steven ; Manning, Keefe B. / A fluid dynamics study in a 50 cc pulsatile ventricular assist device : Influence of heart rate variability. In: Journal of Biomechanical Engineering. 2011 ; Vol. 133, No. 10.
@article{8b9472a309ff4b2498d5d3b09c9ac6f0,
title = "A fluid dynamics study in a 50 cc pulsatile ventricular assist device: Influence of heart rate variability",
abstract = "Although left ventricular assist devices (LVADs) have had success in supporting severe heart failure patients, thrombus formation within these devices still limits their long term use. Research has shown that thrombosis in the Penn State pulsatile LVAD, on a polyurethane blood sac, is largely a function of the underlying fluid mechanics and may be correlated to wall shear rates below 500 s -1. Given the large range of heart rate and systolic durations employed, in vivo it is useful to study the fluid mechanics of pulsatile LVADs under these conditions. Particle image velocimetry (PIV) was used to capture planar flow in the pump body of a Penn State 50 cubic centimeters (cc) LVAD for heart rates of 75-150 bpm and respective systolic durations of 38-50. Shear rates were calculated along the lower device wall with attention given to the uncertainty of the shear rate measurement as a function of pixel magnification. Spatial and temporal shear rate changes associated with data collection frequency were also investigated. The accuracy of the shear rate calculation improved by approximately 40 as the resolution increased from 35 to 12 m/pixel. In addition, data collection in 10 ms, rather than 50 ms, intervals was found to be preferable. Increasing heart rate and systolic duration showed little change in wall shear rate patterns, with wall shear rate magnitude scaling by approximately the kinematic viscosity divided by the square of the average inlet velocity, which is essentially half the friction coefficient. Changes in in vivo operating conditions strongly influence wall shear rates within our device, and likely play a significant role in thrombus deposition. Refinement of PIV techniques at higher magnifications can be useful in moving towards better prediction of thrombosis in LVADs.",
author = "Nanna, {Jason C.} and Navitsky, {Michael A.} and Topper, {Stephen R.} and Steven Deutsch and Manning, {Keefe B.}",
year = "2011",
month = "11",
day = "10",
doi = "10.1115/1.4005001",
language = "English (US)",
volume = "133",
journal = "Journal of Biomechanical Engineering",
issn = "0148-0731",
publisher = "American Society of Mechanical Engineers(ASME)",
number = "10",

}

A fluid dynamics study in a 50 cc pulsatile ventricular assist device : Influence of heart rate variability. / Nanna, Jason C.; Navitsky, Michael A.; Topper, Stephen R.; Deutsch, Steven; Manning, Keefe B.

In: Journal of Biomechanical Engineering, Vol. 133, No. 10, 101002, 10.11.2011.

Research output: Contribution to journalArticle

TY - JOUR

T1 - A fluid dynamics study in a 50 cc pulsatile ventricular assist device

T2 - Influence of heart rate variability

AU - Nanna, Jason C.

AU - Navitsky, Michael A.

AU - Topper, Stephen R.

AU - Deutsch, Steven

AU - Manning, Keefe B.

PY - 2011/11/10

Y1 - 2011/11/10

N2 - Although left ventricular assist devices (LVADs) have had success in supporting severe heart failure patients, thrombus formation within these devices still limits their long term use. Research has shown that thrombosis in the Penn State pulsatile LVAD, on a polyurethane blood sac, is largely a function of the underlying fluid mechanics and may be correlated to wall shear rates below 500 s -1. Given the large range of heart rate and systolic durations employed, in vivo it is useful to study the fluid mechanics of pulsatile LVADs under these conditions. Particle image velocimetry (PIV) was used to capture planar flow in the pump body of a Penn State 50 cubic centimeters (cc) LVAD for heart rates of 75-150 bpm and respective systolic durations of 38-50. Shear rates were calculated along the lower device wall with attention given to the uncertainty of the shear rate measurement as a function of pixel magnification. Spatial and temporal shear rate changes associated with data collection frequency were also investigated. The accuracy of the shear rate calculation improved by approximately 40 as the resolution increased from 35 to 12 m/pixel. In addition, data collection in 10 ms, rather than 50 ms, intervals was found to be preferable. Increasing heart rate and systolic duration showed little change in wall shear rate patterns, with wall shear rate magnitude scaling by approximately the kinematic viscosity divided by the square of the average inlet velocity, which is essentially half the friction coefficient. Changes in in vivo operating conditions strongly influence wall shear rates within our device, and likely play a significant role in thrombus deposition. Refinement of PIV techniques at higher magnifications can be useful in moving towards better prediction of thrombosis in LVADs.

AB - Although left ventricular assist devices (LVADs) have had success in supporting severe heart failure patients, thrombus formation within these devices still limits their long term use. Research has shown that thrombosis in the Penn State pulsatile LVAD, on a polyurethane blood sac, is largely a function of the underlying fluid mechanics and may be correlated to wall shear rates below 500 s -1. Given the large range of heart rate and systolic durations employed, in vivo it is useful to study the fluid mechanics of pulsatile LVADs under these conditions. Particle image velocimetry (PIV) was used to capture planar flow in the pump body of a Penn State 50 cubic centimeters (cc) LVAD for heart rates of 75-150 bpm and respective systolic durations of 38-50. Shear rates were calculated along the lower device wall with attention given to the uncertainty of the shear rate measurement as a function of pixel magnification. Spatial and temporal shear rate changes associated with data collection frequency were also investigated. The accuracy of the shear rate calculation improved by approximately 40 as the resolution increased from 35 to 12 m/pixel. In addition, data collection in 10 ms, rather than 50 ms, intervals was found to be preferable. Increasing heart rate and systolic duration showed little change in wall shear rate patterns, with wall shear rate magnitude scaling by approximately the kinematic viscosity divided by the square of the average inlet velocity, which is essentially half the friction coefficient. Changes in in vivo operating conditions strongly influence wall shear rates within our device, and likely play a significant role in thrombus deposition. Refinement of PIV techniques at higher magnifications can be useful in moving towards better prediction of thrombosis in LVADs.

UR - http://www.scopus.com/inward/record.url?scp=80555133150&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=80555133150&partnerID=8YFLogxK

U2 - 10.1115/1.4005001

DO - 10.1115/1.4005001

M3 - Article

C2 - 22070327

AN - SCOPUS:80555133150

VL - 133

JO - Journal of Biomechanical Engineering

JF - Journal of Biomechanical Engineering

SN - 0148-0731

IS - 10

M1 - 101002

ER -