Advanced flow-through external pressure-balanced reference electrode for potentiometric and pH studies in high temperature aqueous solutions

Serguei Lvov, H. Gao, D. D. Macdonald

Research output: Contribution to journalArticle

32 Citations (Scopus)

Abstract

A new flow-through external pressure-balanced reference electrode (FTEPBRE) has been developed for potentiometric and pH measurements in high temperature aqueous solutions. The unique feature of this advanced FTEPBRE is that the reference solution flows through the electrode so that the concentration of solution across the thermal liquid junction is well defined. Because the electrolyte concentration profile in the reference electrode is maintained constant, uncertainty in the thermal liquid junction potential (TLJP) can be eliminated at a given temperature and pressure. We have employed the silver|silver chloride|platinum|hydrogen non-isothermal electrochemical system for potentiometric measurements of potentials between HCl(aq) solutions of 0.01 and 0.001 mol kg-1 at temperatures up to 633 K and at pressures of 275 and 338 bar. The results of the measurements have been compared with calculations made using the available thermodynamic data, and good agreement, within ± 3 to 10 mV, between the calculated and the measured potentials is obtained. We concluded that the advanced reference electrode can be used to make accurate potentiometric measurements (within a few mV) in aqueous systems over wide ranges of temperatures and pressures. Also we demonstrated an ability of the thermocell to measure pH with high accuracy of ±0.04 to 0.09 over wide ranges of temperature and pressure.

Original languageEnglish (US)
Pages (from-to)186-194
Number of pages9
JournalJournal of Electroanalytical Chemistry
Volume443
Issue number2
DOIs
StatePublished - Feb 20 1998

Fingerprint

Electrodes
Temperature
Silver
Liquids
Platinum
Electrolytes
Hydrogen
Thermodynamics
Hot Temperature

All Science Journal Classification (ASJC) codes

  • Analytical Chemistry
  • Chemical Engineering(all)
  • Electrochemistry

Cite this

@article{958e5c57b3d74f08897867f761eb2d00,
title = "Advanced flow-through external pressure-balanced reference electrode for potentiometric and pH studies in high temperature aqueous solutions",
abstract = "A new flow-through external pressure-balanced reference electrode (FTEPBRE) has been developed for potentiometric and pH measurements in high temperature aqueous solutions. The unique feature of this advanced FTEPBRE is that the reference solution flows through the electrode so that the concentration of solution across the thermal liquid junction is well defined. Because the electrolyte concentration profile in the reference electrode is maintained constant, uncertainty in the thermal liquid junction potential (TLJP) can be eliminated at a given temperature and pressure. We have employed the silver|silver chloride|platinum|hydrogen non-isothermal electrochemical system for potentiometric measurements of potentials between HCl(aq) solutions of 0.01 and 0.001 mol kg-1 at temperatures up to 633 K and at pressures of 275 and 338 bar. The results of the measurements have been compared with calculations made using the available thermodynamic data, and good agreement, within ± 3 to 10 mV, between the calculated and the measured potentials is obtained. We concluded that the advanced reference electrode can be used to make accurate potentiometric measurements (within a few mV) in aqueous systems over wide ranges of temperatures and pressures. Also we demonstrated an ability of the thermocell to measure pH with high accuracy of ±0.04 to 0.09 over wide ranges of temperature and pressure.",
author = "Serguei Lvov and H. Gao and Macdonald, {D. D.}",
year = "1998",
month = "2",
day = "20",
doi = "10.1016/S0022-0728(97)00530-5",
language = "English (US)",
volume = "443",
pages = "186--194",
journal = "Journal of Electroanalytical Chemistry",
issn = "0368-1874",
publisher = "Elsevier Sequoia",
number = "2",

}

Advanced flow-through external pressure-balanced reference electrode for potentiometric and pH studies in high temperature aqueous solutions. / Lvov, Serguei; Gao, H.; Macdonald, D. D.

In: Journal of Electroanalytical Chemistry, Vol. 443, No. 2, 20.02.1998, p. 186-194.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Advanced flow-through external pressure-balanced reference electrode for potentiometric and pH studies in high temperature aqueous solutions

AU - Lvov, Serguei

AU - Gao, H.

AU - Macdonald, D. D.

PY - 1998/2/20

Y1 - 1998/2/20

N2 - A new flow-through external pressure-balanced reference electrode (FTEPBRE) has been developed for potentiometric and pH measurements in high temperature aqueous solutions. The unique feature of this advanced FTEPBRE is that the reference solution flows through the electrode so that the concentration of solution across the thermal liquid junction is well defined. Because the electrolyte concentration profile in the reference electrode is maintained constant, uncertainty in the thermal liquid junction potential (TLJP) can be eliminated at a given temperature and pressure. We have employed the silver|silver chloride|platinum|hydrogen non-isothermal electrochemical system for potentiometric measurements of potentials between HCl(aq) solutions of 0.01 and 0.001 mol kg-1 at temperatures up to 633 K and at pressures of 275 and 338 bar. The results of the measurements have been compared with calculations made using the available thermodynamic data, and good agreement, within ± 3 to 10 mV, between the calculated and the measured potentials is obtained. We concluded that the advanced reference electrode can be used to make accurate potentiometric measurements (within a few mV) in aqueous systems over wide ranges of temperatures and pressures. Also we demonstrated an ability of the thermocell to measure pH with high accuracy of ±0.04 to 0.09 over wide ranges of temperature and pressure.

AB - A new flow-through external pressure-balanced reference electrode (FTEPBRE) has been developed for potentiometric and pH measurements in high temperature aqueous solutions. The unique feature of this advanced FTEPBRE is that the reference solution flows through the electrode so that the concentration of solution across the thermal liquid junction is well defined. Because the electrolyte concentration profile in the reference electrode is maintained constant, uncertainty in the thermal liquid junction potential (TLJP) can be eliminated at a given temperature and pressure. We have employed the silver|silver chloride|platinum|hydrogen non-isothermal electrochemical system for potentiometric measurements of potentials between HCl(aq) solutions of 0.01 and 0.001 mol kg-1 at temperatures up to 633 K and at pressures of 275 and 338 bar. The results of the measurements have been compared with calculations made using the available thermodynamic data, and good agreement, within ± 3 to 10 mV, between the calculated and the measured potentials is obtained. We concluded that the advanced reference electrode can be used to make accurate potentiometric measurements (within a few mV) in aqueous systems over wide ranges of temperatures and pressures. Also we demonstrated an ability of the thermocell to measure pH with high accuracy of ±0.04 to 0.09 over wide ranges of temperature and pressure.

UR - http://www.scopus.com/inward/record.url?scp=0032003877&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0032003877&partnerID=8YFLogxK

U2 - 10.1016/S0022-0728(97)00530-5

DO - 10.1016/S0022-0728(97)00530-5

M3 - Article

AN - SCOPUS:0032003877

VL - 443

SP - 186

EP - 194

JO - Journal of Electroanalytical Chemistry

JF - Journal of Electroanalytical Chemistry

SN - 0368-1874

IS - 2

ER -