An evaluation of a handheld indirect calorimeter against a standard calorimeter in obese and nonobese adults

David Frankenfield, Abigail Coleman

Research output: Contribution to journalArticle

6 Citations (Scopus)

Abstract

Background: Handheld indirect calorimetry has the potential to allow simple and inexpensive measurement of resting metabolic rate in spontaneously breathing people. However validation work on these devices is contradictory. The purpose of the current study was to determine the bias and level of agreement of oxygen consumption and resting metabolic rate as measured by a handheld indirect calorimeter against a standard open-circuit indirect calorimetry cart. Materials and Methods: One hundred community-living, spontaneously breathing, ambulatory nonobese and obese adults were studied in single sessions by a single investigator. Sequential measurements were undertaken using the handheld indirect calorimeter and the standard metabolic cart. Measurement sequence was varied randomly. Results: The mean value for oxygen consumption and metabolic rate of the 2 devices was not significantly different. However, agreement between the 2 devices was only 43% in nonobese and obese participants, and there was proportional and fixed bias, with the handheld calorimeter tending to produce a higher value for oxygen consumption and resting metabolic rate. Limits of agreement for resting metabolic rate between the 2 calorimeters were -240 to +300 kcal/d. Conclusions: Measurements of resting metabolic rate by the handheld indirect calorimeter tested in this study are not equivalent to measurements by standard indirect calorimetry.

Original languageEnglish (US)
Pages (from-to)652-658
Number of pages7
JournalJournal of Parenteral and Enteral Nutrition
Volume37
Issue number5
DOIs
StatePublished - Sep 1 2013

Fingerprint

Basal Metabolism
Indirect Calorimetry
Oxygen Consumption
Equipment and Supplies
Respiration
Research Personnel

All Science Journal Classification (ASJC) codes

  • Medicine (miscellaneous)
  • Nutrition and Dietetics

Cite this

@article{0839626511aa4dc5ae92b9bbd34e386b,
title = "An evaluation of a handheld indirect calorimeter against a standard calorimeter in obese and nonobese adults",
abstract = "Background: Handheld indirect calorimetry has the potential to allow simple and inexpensive measurement of resting metabolic rate in spontaneously breathing people. However validation work on these devices is contradictory. The purpose of the current study was to determine the bias and level of agreement of oxygen consumption and resting metabolic rate as measured by a handheld indirect calorimeter against a standard open-circuit indirect calorimetry cart. Materials and Methods: One hundred community-living, spontaneously breathing, ambulatory nonobese and obese adults were studied in single sessions by a single investigator. Sequential measurements were undertaken using the handheld indirect calorimeter and the standard metabolic cart. Measurement sequence was varied randomly. Results: The mean value for oxygen consumption and metabolic rate of the 2 devices was not significantly different. However, agreement between the 2 devices was only 43{\%} in nonobese and obese participants, and there was proportional and fixed bias, with the handheld calorimeter tending to produce a higher value for oxygen consumption and resting metabolic rate. Limits of agreement for resting metabolic rate between the 2 calorimeters were -240 to +300 kcal/d. Conclusions: Measurements of resting metabolic rate by the handheld indirect calorimeter tested in this study are not equivalent to measurements by standard indirect calorimetry.",
author = "David Frankenfield and Abigail Coleman",
year = "2013",
month = "9",
day = "1",
doi = "10.1177/0148607112473340",
language = "English (US)",
volume = "37",
pages = "652--658",
journal = "Journal of Parenteral and Enteral Nutrition",
issn = "0148-6071",
publisher = "SAGE Publications Inc.",
number = "5",

}

An evaluation of a handheld indirect calorimeter against a standard calorimeter in obese and nonobese adults. / Frankenfield, David; Coleman, Abigail.

In: Journal of Parenteral and Enteral Nutrition, Vol. 37, No. 5, 01.09.2013, p. 652-658.

Research output: Contribution to journalArticle

TY - JOUR

T1 - An evaluation of a handheld indirect calorimeter against a standard calorimeter in obese and nonobese adults

AU - Frankenfield, David

AU - Coleman, Abigail

PY - 2013/9/1

Y1 - 2013/9/1

N2 - Background: Handheld indirect calorimetry has the potential to allow simple and inexpensive measurement of resting metabolic rate in spontaneously breathing people. However validation work on these devices is contradictory. The purpose of the current study was to determine the bias and level of agreement of oxygen consumption and resting metabolic rate as measured by a handheld indirect calorimeter against a standard open-circuit indirect calorimetry cart. Materials and Methods: One hundred community-living, spontaneously breathing, ambulatory nonobese and obese adults were studied in single sessions by a single investigator. Sequential measurements were undertaken using the handheld indirect calorimeter and the standard metabolic cart. Measurement sequence was varied randomly. Results: The mean value for oxygen consumption and metabolic rate of the 2 devices was not significantly different. However, agreement between the 2 devices was only 43% in nonobese and obese participants, and there was proportional and fixed bias, with the handheld calorimeter tending to produce a higher value for oxygen consumption and resting metabolic rate. Limits of agreement for resting metabolic rate between the 2 calorimeters were -240 to +300 kcal/d. Conclusions: Measurements of resting metabolic rate by the handheld indirect calorimeter tested in this study are not equivalent to measurements by standard indirect calorimetry.

AB - Background: Handheld indirect calorimetry has the potential to allow simple and inexpensive measurement of resting metabolic rate in spontaneously breathing people. However validation work on these devices is contradictory. The purpose of the current study was to determine the bias and level of agreement of oxygen consumption and resting metabolic rate as measured by a handheld indirect calorimeter against a standard open-circuit indirect calorimetry cart. Materials and Methods: One hundred community-living, spontaneously breathing, ambulatory nonobese and obese adults were studied in single sessions by a single investigator. Sequential measurements were undertaken using the handheld indirect calorimeter and the standard metabolic cart. Measurement sequence was varied randomly. Results: The mean value for oxygen consumption and metabolic rate of the 2 devices was not significantly different. However, agreement between the 2 devices was only 43% in nonobese and obese participants, and there was proportional and fixed bias, with the handheld calorimeter tending to produce a higher value for oxygen consumption and resting metabolic rate. Limits of agreement for resting metabolic rate between the 2 calorimeters were -240 to +300 kcal/d. Conclusions: Measurements of resting metabolic rate by the handheld indirect calorimeter tested in this study are not equivalent to measurements by standard indirect calorimetry.

UR - http://www.scopus.com/inward/record.url?scp=84883521690&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84883521690&partnerID=8YFLogxK

U2 - 10.1177/0148607112473340

DO - 10.1177/0148607112473340

M3 - Article

C2 - 23334358

AN - SCOPUS:84883521690

VL - 37

SP - 652

EP - 658

JO - Journal of Parenteral and Enteral Nutrition

JF - Journal of Parenteral and Enteral Nutrition

SN - 0148-6071

IS - 5

ER -