Ethylene oxide (EO) gas is widely used in hospitals to sterilize certain moisture- and heat-sensitive materials. Based on scientific studies indicating its potential as a human carcinogen and mutagen, and the possible genotoxic, reproductive, neurologic, and sensitization hazards associated with EO exposure, the Occupational Safety and Health Administration (OSHA) has recently lowered the permissible exposure limit (PEL) from 50 ppm to 1 ppm as an 8-hour time-weighted average (TWA). This standard also established an 'action level' of 0. 5 ppm for an 8-hour TWA, below which employers are exempted from such requirements as periodic employee exposure monitoring or medical surveillance. The capabilities of five EO diffusion monitors were examined in the TWA concentration range of 0. 25-3. 7 ppm. Both accuracy and precision were tested by exposing these devices simultaneously to measured concentrations of EO in a stainless steel exposure chamber. Temperature and humidity conditions were controlled, as was the flow rate of the gases across the sampling areas of the diffusion monitors.

Original languageEnglish (US)
Pages (from-to)309-317
Number of pages9
JournalMedical Instrumentation
Volume18
Issue number6
StatePublished - Nov 1 1984

Fingerprint

Ethylene Oxide
Ethylene
Oxides
Gases
Mutagens
United States Occupational Safety and Health Administration
Carcinogens
Stainless Steel
Humidity
Nervous System
Atmospheric humidity
Hazards
Moisture
Stainless steel
Hot Temperature
Flow rate
Health
Personnel
Sampling
Equipment and Supplies

All Science Journal Classification (ASJC) codes

  • Bioengineering
  • Medicine (miscellaneous)

Cite this

@article{afd22d2463e743bda5fa528b373e9134,
title = "An evaluation of the accuracy and reliability of ethylene oxide diffusion badge monitors",
abstract = "Ethylene oxide (EO) gas is widely used in hospitals to sterilize certain moisture- and heat-sensitive materials. Based on scientific studies indicating its potential as a human carcinogen and mutagen, and the possible genotoxic, reproductive, neurologic, and sensitization hazards associated with EO exposure, the Occupational Safety and Health Administration (OSHA) has recently lowered the permissible exposure limit (PEL) from 50 ppm to 1 ppm as an 8-hour time-weighted average (TWA). This standard also established an 'action level' of 0. 5 ppm for an 8-hour TWA, below which employers are exempted from such requirements as periodic employee exposure monitoring or medical surveillance. The capabilities of five EO diffusion monitors were examined in the TWA concentration range of 0. 25-3. 7 ppm. Both accuracy and precision were tested by exposing these devices simultaneously to measured concentrations of EO in a stainless steel exposure chamber. Temperature and humidity conditions were controlled, as was the flow rate of the gases across the sampling areas of the diffusion monitors.",
author = "Loving, {T. J.} and Allen, {R. O.} and Mosher, {T. J.}",
year = "1984",
month = "11",
day = "1",
language = "English (US)",
volume = "18",
pages = "309--317",
journal = "Biomedical Instrumentation and Technology",
issn = "0899-8205",
publisher = "Association for the Advancement of Medical Instrumentation",
number = "6",

}

An evaluation of the accuracy and reliability of ethylene oxide diffusion badge monitors. / Loving, T. J.; Allen, R. O.; Mosher, T. J.

In: Medical Instrumentation, Vol. 18, No. 6, 01.11.1984, p. 309-317.

Research output: Contribution to journalArticle

TY - JOUR

T1 - An evaluation of the accuracy and reliability of ethylene oxide diffusion badge monitors

AU - Loving, T. J.

AU - Allen, R. O.

AU - Mosher, T. J.

PY - 1984/11/1

Y1 - 1984/11/1

N2 - Ethylene oxide (EO) gas is widely used in hospitals to sterilize certain moisture- and heat-sensitive materials. Based on scientific studies indicating its potential as a human carcinogen and mutagen, and the possible genotoxic, reproductive, neurologic, and sensitization hazards associated with EO exposure, the Occupational Safety and Health Administration (OSHA) has recently lowered the permissible exposure limit (PEL) from 50 ppm to 1 ppm as an 8-hour time-weighted average (TWA). This standard also established an 'action level' of 0. 5 ppm for an 8-hour TWA, below which employers are exempted from such requirements as periodic employee exposure monitoring or medical surveillance. The capabilities of five EO diffusion monitors were examined in the TWA concentration range of 0. 25-3. 7 ppm. Both accuracy and precision were tested by exposing these devices simultaneously to measured concentrations of EO in a stainless steel exposure chamber. Temperature and humidity conditions were controlled, as was the flow rate of the gases across the sampling areas of the diffusion monitors.

AB - Ethylene oxide (EO) gas is widely used in hospitals to sterilize certain moisture- and heat-sensitive materials. Based on scientific studies indicating its potential as a human carcinogen and mutagen, and the possible genotoxic, reproductive, neurologic, and sensitization hazards associated with EO exposure, the Occupational Safety and Health Administration (OSHA) has recently lowered the permissible exposure limit (PEL) from 50 ppm to 1 ppm as an 8-hour time-weighted average (TWA). This standard also established an 'action level' of 0. 5 ppm for an 8-hour TWA, below which employers are exempted from such requirements as periodic employee exposure monitoring or medical surveillance. The capabilities of five EO diffusion monitors were examined in the TWA concentration range of 0. 25-3. 7 ppm. Both accuracy and precision were tested by exposing these devices simultaneously to measured concentrations of EO in a stainless steel exposure chamber. Temperature and humidity conditions were controlled, as was the flow rate of the gases across the sampling areas of the diffusion monitors.

UR - http://www.scopus.com/inward/record.url?scp=0021531756&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0021531756&partnerID=8YFLogxK

M3 - Article

C2 - 6521676

AN - SCOPUS:0021531756

VL - 18

SP - 309

EP - 317

JO - Biomedical Instrumentation and Technology

JF - Biomedical Instrumentation and Technology

SN - 0899-8205

IS - 6

ER -