Anatomic and biomechanical assessment of transarticular screw fixation for atlantoaxial instability

Peter B. Hanson, Pasquale X. Montesano, Neil A. Sharkey, Wolfgang Rauschning

Research output: Contribution to journalArticle

114 Citations (Scopus)

Abstract

The purpose of this study is to elucidate anatomically the atlantoaxial transarticular screw fixation described by Magerl in 1979 and compare it biomechanically with Gallie wiring. Five human C1-C2 specimens were tested in flexion/extension and rotation intact, then after wiring and screw fixation. Mean screw length was 39 mm, 25 mm in the C2 lamina and 14 mm in the lateral mass. Angular displacement of screwed specimens was significantly less than control or wired groups. Stiffness at 0-0.5 Nm loads was significantly greater for screwed specimens than for wired or controls (101 ± 49 Nm, 10.3 ± 9.2 Nm, and 1.96 ± 0.18 Nm, respectively). All specimens withstood 5 Nm in flexion and extension without failure. Screw fixation provides stability comparable to Gallie wiring and is stiffer at low-range forces and rotational angles.

Original languageEnglish (US)
Pages (from-to)1141-1145
Number of pages5
JournalSpine
Volume16
Issue number10
DOIs
StatePublished - Oct 1991

All Science Journal Classification (ASJC) codes

  • Orthopedics and Sports Medicine
  • Clinical Neurology

Cite this

Hanson, Peter B. ; Montesano, Pasquale X. ; Sharkey, Neil A. ; Rauschning, Wolfgang. / Anatomic and biomechanical assessment of transarticular screw fixation for atlantoaxial instability. In: Spine. 1991 ; Vol. 16, No. 10. pp. 1141-1145.
@article{9598a50e65a846c699a4f065b283a619,
title = "Anatomic and biomechanical assessment of transarticular screw fixation for atlantoaxial instability",
abstract = "The purpose of this study is to elucidate anatomically the atlantoaxial transarticular screw fixation described by Magerl in 1979 and compare it biomechanically with Gallie wiring. Five human C1-C2 specimens were tested in flexion/extension and rotation intact, then after wiring and screw fixation. Mean screw length was 39 mm, 25 mm in the C2 lamina and 14 mm in the lateral mass. Angular displacement of screwed specimens was significantly less than control or wired groups. Stiffness at 0-0.5 Nm loads was significantly greater for screwed specimens than for wired or controls (101 ± 49 Nm, 10.3 ± 9.2 Nm, and 1.96 ± 0.18 Nm, respectively). All specimens withstood 5 Nm in flexion and extension without failure. Screw fixation provides stability comparable to Gallie wiring and is stiffer at low-range forces and rotational angles.",
author = "Hanson, {Peter B.} and Montesano, {Pasquale X.} and Sharkey, {Neil A.} and Wolfgang Rauschning",
year = "1991",
month = "10",
doi = "10.1097/00007632-199110000-00002",
language = "English (US)",
volume = "16",
pages = "1141--1145",
journal = "Spine",
issn = "0362-2436",
publisher = "Lippincott Williams and Wilkins",
number = "10",

}

Anatomic and biomechanical assessment of transarticular screw fixation for atlantoaxial instability. / Hanson, Peter B.; Montesano, Pasquale X.; Sharkey, Neil A.; Rauschning, Wolfgang.

In: Spine, Vol. 16, No. 10, 10.1991, p. 1141-1145.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Anatomic and biomechanical assessment of transarticular screw fixation for atlantoaxial instability

AU - Hanson, Peter B.

AU - Montesano, Pasquale X.

AU - Sharkey, Neil A.

AU - Rauschning, Wolfgang

PY - 1991/10

Y1 - 1991/10

N2 - The purpose of this study is to elucidate anatomically the atlantoaxial transarticular screw fixation described by Magerl in 1979 and compare it biomechanically with Gallie wiring. Five human C1-C2 specimens were tested in flexion/extension and rotation intact, then after wiring and screw fixation. Mean screw length was 39 mm, 25 mm in the C2 lamina and 14 mm in the lateral mass. Angular displacement of screwed specimens was significantly less than control or wired groups. Stiffness at 0-0.5 Nm loads was significantly greater for screwed specimens than for wired or controls (101 ± 49 Nm, 10.3 ± 9.2 Nm, and 1.96 ± 0.18 Nm, respectively). All specimens withstood 5 Nm in flexion and extension without failure. Screw fixation provides stability comparable to Gallie wiring and is stiffer at low-range forces and rotational angles.

AB - The purpose of this study is to elucidate anatomically the atlantoaxial transarticular screw fixation described by Magerl in 1979 and compare it biomechanically with Gallie wiring. Five human C1-C2 specimens were tested in flexion/extension and rotation intact, then after wiring and screw fixation. Mean screw length was 39 mm, 25 mm in the C2 lamina and 14 mm in the lateral mass. Angular displacement of screwed specimens was significantly less than control or wired groups. Stiffness at 0-0.5 Nm loads was significantly greater for screwed specimens than for wired or controls (101 ± 49 Nm, 10.3 ± 9.2 Nm, and 1.96 ± 0.18 Nm, respectively). All specimens withstood 5 Nm in flexion and extension without failure. Screw fixation provides stability comparable to Gallie wiring and is stiffer at low-range forces and rotational angles.

UR - http://www.scopus.com/inward/record.url?scp=0025998726&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0025998726&partnerID=8YFLogxK

U2 - 10.1097/00007632-199110000-00002

DO - 10.1097/00007632-199110000-00002

M3 - Article

C2 - 1754933

AN - SCOPUS:0025998726

VL - 16

SP - 1141

EP - 1145

JO - Spine

JF - Spine

SN - 0362-2436

IS - 10

ER -