Characterizing the HeartMate II left ventricular assist device outflow using particle image velocimetry

Grant W. Rowlands, Bryan C. Good, Steven Deutsch, Keefe B. Manning

Research output: Contribution to journalArticle

1 Citation (Scopus)

Abstract

Ventricular assist devices (VADs) are implanted in patients with a diseased ventricle to maintain peripheral perfusion as a bridge-to-transplant or as destination therapy. However, some patients with continuous flow VADs (e.g., HeartMate II (HMII)) have experienced gastrointestinal (GI) bleeding, in part caused by the proteolytic cleavage or mechanical destruction of von Willebrand factor (vWF), a clotting glycoprotein. in vitro studies were performed to measure the flow located within the HMII outlet cannula under both steady and physiological conditions using particle image velocimetry (PIV). Under steady flow, a mock flow loop was used with the HMII producing a flow rate of 3.2 L/min. The physiological experiment included a pulsatile pump operated at 105 BPM with a ventricle filling volume of 50 mL and in conjunction with the HMII producing a total flow rate of 5.0 L/min. Velocity fields, Reynolds normal stresses (RNSs), and Reynolds shear stresses (RSSs) were analyzed to quantify the outlet flow's potential contribution to vWF degradation. Under both flow conditions, the HMII generated principal Reynolds stresses that are, at times, orders of magnitude higher than those needed to unfurl vWF, potentially impacting its physiological function. Under steady flow, principal RNSs were calculated to be approximately 500 Pa in the outlet cannula. Elevated Reynolds stresses were observed throughout every phase of the cardiac cycle under physiological flow with principal RNSs approaching 1500 Pa during peak systole. Prolonged exposure to these conditions may lead to acquired von Willebrand syndrome (AvWS), which is accompanied by uncontrollable bleeding episodes.

Original languageEnglish (US)
Article number071008
JournalJournal of Biomechanical Engineering
Volume140
Issue number7
DOIs
StatePublished - Jul 1 2018

Fingerprint

Left ventricular assist devices
Heart-Assist Devices
Rheology
Velocity measurement
von Willebrand Factor
Steady flow
Hemorrhage
Flow rate
Systole
Transplants
Glycoproteins
Potential flow
Perfusion
Shear stress
Pumps
Degradation
Experiments
Cannula

All Science Journal Classification (ASJC) codes

  • Biomedical Engineering
  • Physiology (medical)

Cite this

@article{55ca10c687a34d4b965931a1a31636f7,
title = "Characterizing the HeartMate II left ventricular assist device outflow using particle image velocimetry",
abstract = "Ventricular assist devices (VADs) are implanted in patients with a diseased ventricle to maintain peripheral perfusion as a bridge-to-transplant or as destination therapy. However, some patients with continuous flow VADs (e.g., HeartMate II (HMII)) have experienced gastrointestinal (GI) bleeding, in part caused by the proteolytic cleavage or mechanical destruction of von Willebrand factor (vWF), a clotting glycoprotein. in vitro studies were performed to measure the flow located within the HMII outlet cannula under both steady and physiological conditions using particle image velocimetry (PIV). Under steady flow, a mock flow loop was used with the HMII producing a flow rate of 3.2 L/min. The physiological experiment included a pulsatile pump operated at 105 BPM with a ventricle filling volume of 50 mL and in conjunction with the HMII producing a total flow rate of 5.0 L/min. Velocity fields, Reynolds normal stresses (RNSs), and Reynolds shear stresses (RSSs) were analyzed to quantify the outlet flow's potential contribution to vWF degradation. Under both flow conditions, the HMII generated principal Reynolds stresses that are, at times, orders of magnitude higher than those needed to unfurl vWF, potentially impacting its physiological function. Under steady flow, principal RNSs were calculated to be approximately 500 Pa in the outlet cannula. Elevated Reynolds stresses were observed throughout every phase of the cardiac cycle under physiological flow with principal RNSs approaching 1500 Pa during peak systole. Prolonged exposure to these conditions may lead to acquired von Willebrand syndrome (AvWS), which is accompanied by uncontrollable bleeding episodes.",
author = "Rowlands, {Grant W.} and Good, {Bryan C.} and Steven Deutsch and Manning, {Keefe B.}",
year = "2018",
month = "7",
day = "1",
doi = "10.1115/1.4039822",
language = "English (US)",
volume = "140",
journal = "Journal of Biomechanical Engineering",
issn = "0148-0731",
publisher = "American Society of Mechanical Engineers(ASME)",
number = "7",

}

Characterizing the HeartMate II left ventricular assist device outflow using particle image velocimetry. / Rowlands, Grant W.; Good, Bryan C.; Deutsch, Steven; Manning, Keefe B.

In: Journal of Biomechanical Engineering, Vol. 140, No. 7, 071008, 01.07.2018.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Characterizing the HeartMate II left ventricular assist device outflow using particle image velocimetry

AU - Rowlands, Grant W.

AU - Good, Bryan C.

AU - Deutsch, Steven

AU - Manning, Keefe B.

PY - 2018/7/1

Y1 - 2018/7/1

N2 - Ventricular assist devices (VADs) are implanted in patients with a diseased ventricle to maintain peripheral perfusion as a bridge-to-transplant or as destination therapy. However, some patients with continuous flow VADs (e.g., HeartMate II (HMII)) have experienced gastrointestinal (GI) bleeding, in part caused by the proteolytic cleavage or mechanical destruction of von Willebrand factor (vWF), a clotting glycoprotein. in vitro studies were performed to measure the flow located within the HMII outlet cannula under both steady and physiological conditions using particle image velocimetry (PIV). Under steady flow, a mock flow loop was used with the HMII producing a flow rate of 3.2 L/min. The physiological experiment included a pulsatile pump operated at 105 BPM with a ventricle filling volume of 50 mL and in conjunction with the HMII producing a total flow rate of 5.0 L/min. Velocity fields, Reynolds normal stresses (RNSs), and Reynolds shear stresses (RSSs) were analyzed to quantify the outlet flow's potential contribution to vWF degradation. Under both flow conditions, the HMII generated principal Reynolds stresses that are, at times, orders of magnitude higher than those needed to unfurl vWF, potentially impacting its physiological function. Under steady flow, principal RNSs were calculated to be approximately 500 Pa in the outlet cannula. Elevated Reynolds stresses were observed throughout every phase of the cardiac cycle under physiological flow with principal RNSs approaching 1500 Pa during peak systole. Prolonged exposure to these conditions may lead to acquired von Willebrand syndrome (AvWS), which is accompanied by uncontrollable bleeding episodes.

AB - Ventricular assist devices (VADs) are implanted in patients with a diseased ventricle to maintain peripheral perfusion as a bridge-to-transplant or as destination therapy. However, some patients with continuous flow VADs (e.g., HeartMate II (HMII)) have experienced gastrointestinal (GI) bleeding, in part caused by the proteolytic cleavage or mechanical destruction of von Willebrand factor (vWF), a clotting glycoprotein. in vitro studies were performed to measure the flow located within the HMII outlet cannula under both steady and physiological conditions using particle image velocimetry (PIV). Under steady flow, a mock flow loop was used with the HMII producing a flow rate of 3.2 L/min. The physiological experiment included a pulsatile pump operated at 105 BPM with a ventricle filling volume of 50 mL and in conjunction with the HMII producing a total flow rate of 5.0 L/min. Velocity fields, Reynolds normal stresses (RNSs), and Reynolds shear stresses (RSSs) were analyzed to quantify the outlet flow's potential contribution to vWF degradation. Under both flow conditions, the HMII generated principal Reynolds stresses that are, at times, orders of magnitude higher than those needed to unfurl vWF, potentially impacting its physiological function. Under steady flow, principal RNSs were calculated to be approximately 500 Pa in the outlet cannula. Elevated Reynolds stresses were observed throughout every phase of the cardiac cycle under physiological flow with principal RNSs approaching 1500 Pa during peak systole. Prolonged exposure to these conditions may lead to acquired von Willebrand syndrome (AvWS), which is accompanied by uncontrollable bleeding episodes.

UR - http://www.scopus.com/inward/record.url?scp=85046662844&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85046662844&partnerID=8YFLogxK

U2 - 10.1115/1.4039822

DO - 10.1115/1.4039822

M3 - Article

C2 - 29715362

AN - SCOPUS:85046662844

VL - 140

JO - Journal of Biomechanical Engineering

JF - Journal of Biomechanical Engineering

SN - 0148-0731

IS - 7

M1 - 071008

ER -