CHRONIC ANIMAL STUDIES WITH A MOTOR-DRIVEN LVAD AND AN IMPLANTED COMPLIANCE CHAMBER.

A. Snyder, Gerson Rosenberg, William Weiss, W. Huang, D. Landis, W. S. Pierce

Research output: Contribution to journalConference article

3 Citations (Scopus)

Abstract

The authors have implanted cam type brushless motor driven ventricular assist pump coupled to a sac type compliance chamber in 8 calves. This sealed system has pumped reliably for over 120 days. The study demonstrated the ability to monitor the system and diagnose soft failures as they develop. All failures have been attributed to machining errors and material or component selection; the motor drive design remains viable, with no technological barriers seen to reliable long-term operation.

Original languageEnglish (US)
Pages (from-to)92-97
Number of pages6
JournalTransactions - American Society for Artificial Internal Organs
Volume30
StatePublished - Dec 1 1984

Fingerprint

Compliance
Animals
Cams
Machining
Pumps

All Science Journal Classification (ASJC) codes

  • Medicine(all)

Cite this

@article{0a69872ce6f348b98572ba8aa70a2b0a,
title = "CHRONIC ANIMAL STUDIES WITH A MOTOR-DRIVEN LVAD AND AN IMPLANTED COMPLIANCE CHAMBER.",
abstract = "The authors have implanted cam type brushless motor driven ventricular assist pump coupled to a sac type compliance chamber in 8 calves. This sealed system has pumped reliably for over 120 days. The study demonstrated the ability to monitor the system and diagnose soft failures as they develop. All failures have been attributed to machining errors and material or component selection; the motor drive design remains viable, with no technological barriers seen to reliable long-term operation.",
author = "A. Snyder and Gerson Rosenberg and William Weiss and W. Huang and D. Landis and Pierce, {W. S.}",
year = "1984",
month = "12",
day = "1",
language = "English (US)",
volume = "30",
pages = "92--97",
journal = "ASAIO Journal",
issn = "1058-2916",
publisher = "Lippincott Williams and Wilkins",

}

CHRONIC ANIMAL STUDIES WITH A MOTOR-DRIVEN LVAD AND AN IMPLANTED COMPLIANCE CHAMBER. / Snyder, A.; Rosenberg, Gerson; Weiss, William; Huang, W.; Landis, D.; Pierce, W. S.

In: Transactions - American Society for Artificial Internal Organs, Vol. 30, 01.12.1984, p. 92-97.

Research output: Contribution to journalConference article

TY - JOUR

T1 - CHRONIC ANIMAL STUDIES WITH A MOTOR-DRIVEN LVAD AND AN IMPLANTED COMPLIANCE CHAMBER.

AU - Snyder, A.

AU - Rosenberg, Gerson

AU - Weiss, William

AU - Huang, W.

AU - Landis, D.

AU - Pierce, W. S.

PY - 1984/12/1

Y1 - 1984/12/1

N2 - The authors have implanted cam type brushless motor driven ventricular assist pump coupled to a sac type compliance chamber in 8 calves. This sealed system has pumped reliably for over 120 days. The study demonstrated the ability to monitor the system and diagnose soft failures as they develop. All failures have been attributed to machining errors and material or component selection; the motor drive design remains viable, with no technological barriers seen to reliable long-term operation.

AB - The authors have implanted cam type brushless motor driven ventricular assist pump coupled to a sac type compliance chamber in 8 calves. This sealed system has pumped reliably for over 120 days. The study demonstrated the ability to monitor the system and diagnose soft failures as they develop. All failures have been attributed to machining errors and material or component selection; the motor drive design remains viable, with no technological barriers seen to reliable long-term operation.

UR - http://www.scopus.com/inward/record.url?scp=0021645597&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0021645597&partnerID=8YFLogxK

M3 - Conference article

C2 - 6533976

AN - SCOPUS:0021645597

VL - 30

SP - 92

EP - 97

JO - ASAIO Journal

JF - ASAIO Journal

SN - 1058-2916

ER -