Chronic In Vivo Test of a Right Heart Replacement Blood Pump for Failed Fontan Circulation

Joshua Cysyk, Joseph Brian Clark, Ray Newswanger, Choon-Sik Jhun, Jenelle Izer, Heidi Finicle, John Reibson, Bradley Doxtater, William Weiss, Gerson Rosenberg

Research output: Contribution to journalArticle

1 Citation (Scopus)

Abstract

An implantable rotary blood pump was developed to provide long-term mechanical right heart support for patients who have failing Fontan circulation. The objective of this study was to evaluate the pump in vivo in a 30 day sheep study. Pump speed was set at 3,900 rpm for the duration of the study, and pump power was between 4.3 and 4.6 W. The pump inlet pressures for the superior vena cava (SVC) and inferior vena cava (IVC) were 14 ± 15 and 11 ± 15 mm Hg, respectively, over the duration of the study. Hematocrit remained stable at 30% ± 4%. Partial thromboplastin time (PTT) steadily increased from 30 s preoperatively to a high of 59 s on postoperative day 20, while prothrombin time (PT) remained at 20 ± 2 s for the duration of the study. The implantation and postoperative recovery were successful, and the animal demonstrated normal physiologic pulmonary and venous pressures and cardiac output. On pump inspection, the IVC and SVC inlets were completely clear of any deposits, but there were small thrombi (approximately 0.5 mm diameter) between each of the three rotor blades and along 20% of the parting line of the two volute halves. A complete right heart bypass was performed, postoperative recovery was successful, and the pump demonstrated adequate circulatory support and normal physiologic pulmonary and venous pressures. This study was the first successful test of a right heart replacement device in a chronic animal study.

Original languageEnglish (US)
Pages (from-to)593-600
Number of pages8
JournalASAIO journal (American Society for Artificial Internal Organs : 1992)
Volume65
Issue number6
DOIs
StatePublished - Aug 1 2019

Fingerprint

Superior Vena Cava
Venous Pressure
Inferior Vena Cava
Blood
Right Heart Bypass
Pumps
Lung
Partial Thromboplastin Time
Prothrombin Time
Hematocrit
Cardiac Output
Sheep
Thrombosis
Pressure
Equipment and Supplies
Animals
Recovery
Thromboplastin
Prothrombin
Turbomachine blades

All Science Journal Classification (ASJC) codes

  • Biophysics
  • Bioengineering
  • Biomaterials
  • Biomedical Engineering

Cite this

@article{cd9ecc6a448144b299a04f97956a08d3,
title = "Chronic In Vivo Test of a Right Heart Replacement Blood Pump for Failed Fontan Circulation",
abstract = "An implantable rotary blood pump was developed to provide long-term mechanical right heart support for patients who have failing Fontan circulation. The objective of this study was to evaluate the pump in vivo in a 30 day sheep study. Pump speed was set at 3,900 rpm for the duration of the study, and pump power was between 4.3 and 4.6 W. The pump inlet pressures for the superior vena cava (SVC) and inferior vena cava (IVC) were 14 ± 15 and 11 ± 15 mm Hg, respectively, over the duration of the study. Hematocrit remained stable at 30{\%} ± 4{\%}. Partial thromboplastin time (PTT) steadily increased from 30 s preoperatively to a high of 59 s on postoperative day 20, while prothrombin time (PT) remained at 20 ± 2 s for the duration of the study. The implantation and postoperative recovery were successful, and the animal demonstrated normal physiologic pulmonary and venous pressures and cardiac output. On pump inspection, the IVC and SVC inlets were completely clear of any deposits, but there were small thrombi (approximately 0.5 mm diameter) between each of the three rotor blades and along 20{\%} of the parting line of the two volute halves. A complete right heart bypass was performed, postoperative recovery was successful, and the pump demonstrated adequate circulatory support and normal physiologic pulmonary and venous pressures. This study was the first successful test of a right heart replacement device in a chronic animal study.",
author = "Joshua Cysyk and Clark, {Joseph Brian} and Ray Newswanger and Choon-Sik Jhun and Jenelle Izer and Heidi Finicle and John Reibson and Bradley Doxtater and William Weiss and Gerson Rosenberg",
year = "2019",
month = "8",
day = "1",
doi = "10.1097/MAT.0000000000000888",
language = "English (US)",
volume = "65",
pages = "593--600",
journal = "ASAIO Journal",
issn = "1058-2916",
publisher = "Lippincott Williams and Wilkins",
number = "6",

}

TY - JOUR

T1 - Chronic In Vivo Test of a Right Heart Replacement Blood Pump for Failed Fontan Circulation

AU - Cysyk, Joshua

AU - Clark, Joseph Brian

AU - Newswanger, Ray

AU - Jhun, Choon-Sik

AU - Izer, Jenelle

AU - Finicle, Heidi

AU - Reibson, John

AU - Doxtater, Bradley

AU - Weiss, William

AU - Rosenberg, Gerson

PY - 2019/8/1

Y1 - 2019/8/1

N2 - An implantable rotary blood pump was developed to provide long-term mechanical right heart support for patients who have failing Fontan circulation. The objective of this study was to evaluate the pump in vivo in a 30 day sheep study. Pump speed was set at 3,900 rpm for the duration of the study, and pump power was between 4.3 and 4.6 W. The pump inlet pressures for the superior vena cava (SVC) and inferior vena cava (IVC) were 14 ± 15 and 11 ± 15 mm Hg, respectively, over the duration of the study. Hematocrit remained stable at 30% ± 4%. Partial thromboplastin time (PTT) steadily increased from 30 s preoperatively to a high of 59 s on postoperative day 20, while prothrombin time (PT) remained at 20 ± 2 s for the duration of the study. The implantation and postoperative recovery were successful, and the animal demonstrated normal physiologic pulmonary and venous pressures and cardiac output. On pump inspection, the IVC and SVC inlets were completely clear of any deposits, but there were small thrombi (approximately 0.5 mm diameter) between each of the three rotor blades and along 20% of the parting line of the two volute halves. A complete right heart bypass was performed, postoperative recovery was successful, and the pump demonstrated adequate circulatory support and normal physiologic pulmonary and venous pressures. This study was the first successful test of a right heart replacement device in a chronic animal study.

AB - An implantable rotary blood pump was developed to provide long-term mechanical right heart support for patients who have failing Fontan circulation. The objective of this study was to evaluate the pump in vivo in a 30 day sheep study. Pump speed was set at 3,900 rpm for the duration of the study, and pump power was between 4.3 and 4.6 W. The pump inlet pressures for the superior vena cava (SVC) and inferior vena cava (IVC) were 14 ± 15 and 11 ± 15 mm Hg, respectively, over the duration of the study. Hematocrit remained stable at 30% ± 4%. Partial thromboplastin time (PTT) steadily increased from 30 s preoperatively to a high of 59 s on postoperative day 20, while prothrombin time (PT) remained at 20 ± 2 s for the duration of the study. The implantation and postoperative recovery were successful, and the animal demonstrated normal physiologic pulmonary and venous pressures and cardiac output. On pump inspection, the IVC and SVC inlets were completely clear of any deposits, but there were small thrombi (approximately 0.5 mm diameter) between each of the three rotor blades and along 20% of the parting line of the two volute halves. A complete right heart bypass was performed, postoperative recovery was successful, and the pump demonstrated adequate circulatory support and normal physiologic pulmonary and venous pressures. This study was the first successful test of a right heart replacement device in a chronic animal study.

UR - http://www.scopus.com/inward/record.url?scp=85068035636&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85068035636&partnerID=8YFLogxK

U2 - 10.1097/MAT.0000000000000888

DO - 10.1097/MAT.0000000000000888

M3 - Article

C2 - 30299303

AN - SCOPUS:85068035636

VL - 65

SP - 593

EP - 600

JO - ASAIO Journal

JF - ASAIO Journal

SN - 1058-2916

IS - 6

ER -