Two commercially available radioimmunoassay kits for aldosterone determination not requiring preliminary chromatographic purification were examined. Aldosterone standard curve solutions prepared from the CIS kit, Nuclear International Corp., (x-axis) and Diagnostic Products kit (y-axis) were assayed with both kits and the results yielded a regression equation of y = 0.834x + 0.08 nmol/ liter. Analytical recovery experiments with aldosterone added to urine and serum showed 120% and 101% recovery with the CIS kit, respectively, and 92% and 92% recovery with the Diagnostic Products kit, respectively. Both kits demonstrated good parallelism with urine and serum. Antibody specificity was tested with six structurally related steroids and each kit showed virtually no crossreactivity at clinically normal serum concentrations. Serum aldosterone values were compared before (y-axis) and after (x-axis) column chromatography on Sephadex LH-20, with the following regression equations: y = 1.070x + 0.092 nmol/liter for the CIS kit and y = 1.023x + 0.093 nmol/liter for the Diagnostic Products kit. Regression equations comparing patient samples between kits, CIS (x-axis) and Diagnostic Products (y-axis), were: y = 0.832x + 0.007 μmol/24 h with urine and y = 0.850x + 0.097 nmol/liter with serum.

Original languageEnglish (US)
Pages (from-to)216-221
Number of pages6
JournalClinical Biochemistry
Volume9
Issue numberC
DOIs
StatePublished - 1976

All Science Journal Classification (ASJC) codes

  • Clinical Biochemistry

Fingerprint

Dive into the research topics of 'Comparison of two radioimmunoassay kits for aldosterone determination'. Together they form a unique fingerprint.

Cite this