Development of phantoms for spiral CT

David J. Goodenough, Joshua R. Levy, Claudia Kasales

Research output: Contribution to journalArticle

9 Citations (Scopus)

Abstract

Purpose: This paper reports on the development of a new phantom for spiral CT. The phantom meets the increased demands on phantom z-axis uniformity in order that objects from the CT slice, immediately above and belOW the CT slice of interest, do not contribute perturbing information to the reconstructed CT slice. Material and methods: The phantom depends on formulation of tissue-like materials that can be cast and produced in both geometric and anthropomorphic shapes with sufficient z-axis length to enable unperturbed CT slices of test objects of interest. These materials are then used to produce a series of test objects of CT image quality including low contrast samples that do not require volume averaging or mixing of solutions, and that can reflect sub-slice thickness test objects and supra-slice thickness test objects. Results: The overall phantom and its individual test objects provides meaningful tests of spiral CT image quality including slice sensitivity, CT number linearity and tests of high and low contrast resolution. Schematic designs and actual CT scans are shown. Conclusion: The new spiral phantom appears to meet the increased demands of spiral CT on phantom design, particularly z-axis length, and requirements for low contrast resolution test objects.

Original languageEnglish (US)
Pages (from-to)247-255
Number of pages9
JournalComputerized Medical Imaging and Graphics
Volume22
Issue number3
DOIs
StatePublished - May 1 1998

Fingerprint

Spiral Computed Tomography
Image quality
Computerized tomography
Schematic diagrams
Tissue

All Science Journal Classification (ASJC) codes

  • Radiological and Ultrasound Technology
  • Radiology Nuclear Medicine and imaging
  • Computer Vision and Pattern Recognition
  • Health Informatics
  • Computer Graphics and Computer-Aided Design

Cite this

Goodenough, David J. ; Levy, Joshua R. ; Kasales, Claudia. / Development of phantoms for spiral CT. In: Computerized Medical Imaging and Graphics. 1998 ; Vol. 22, No. 3. pp. 247-255.
@article{0b32c0bf0e8c4aa094beed5127a16583,
title = "Development of phantoms for spiral CT",
abstract = "Purpose: This paper reports on the development of a new phantom for spiral CT. The phantom meets the increased demands on phantom z-axis uniformity in order that objects from the CT slice, immediately above and belOW the CT slice of interest, do not contribute perturbing information to the reconstructed CT slice. Material and methods: The phantom depends on formulation of tissue-like materials that can be cast and produced in both geometric and anthropomorphic shapes with sufficient z-axis length to enable unperturbed CT slices of test objects of interest. These materials are then used to produce a series of test objects of CT image quality including low contrast samples that do not require volume averaging or mixing of solutions, and that can reflect sub-slice thickness test objects and supra-slice thickness test objects. Results: The overall phantom and its individual test objects provides meaningful tests of spiral CT image quality including slice sensitivity, CT number linearity and tests of high and low contrast resolution. Schematic designs and actual CT scans are shown. Conclusion: The new spiral phantom appears to meet the increased demands of spiral CT on phantom design, particularly z-axis length, and requirements for low contrast resolution test objects.",
author = "Goodenough, {David J.} and Levy, {Joshua R.} and Claudia Kasales",
year = "1998",
month = "5",
day = "1",
doi = "10.1016/S0895-6111(98)00020-2",
language = "English (US)",
volume = "22",
pages = "247--255",
journal = "Computerized Medical Imaging and Graphics",
issn = "0895-6111",
publisher = "Elsevier Limited",
number = "3",

}

Development of phantoms for spiral CT. / Goodenough, David J.; Levy, Joshua R.; Kasales, Claudia.

In: Computerized Medical Imaging and Graphics, Vol. 22, No. 3, 01.05.1998, p. 247-255.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Development of phantoms for spiral CT

AU - Goodenough, David J.

AU - Levy, Joshua R.

AU - Kasales, Claudia

PY - 1998/5/1

Y1 - 1998/5/1

N2 - Purpose: This paper reports on the development of a new phantom for spiral CT. The phantom meets the increased demands on phantom z-axis uniformity in order that objects from the CT slice, immediately above and belOW the CT slice of interest, do not contribute perturbing information to the reconstructed CT slice. Material and methods: The phantom depends on formulation of tissue-like materials that can be cast and produced in both geometric and anthropomorphic shapes with sufficient z-axis length to enable unperturbed CT slices of test objects of interest. These materials are then used to produce a series of test objects of CT image quality including low contrast samples that do not require volume averaging or mixing of solutions, and that can reflect sub-slice thickness test objects and supra-slice thickness test objects. Results: The overall phantom and its individual test objects provides meaningful tests of spiral CT image quality including slice sensitivity, CT number linearity and tests of high and low contrast resolution. Schematic designs and actual CT scans are shown. Conclusion: The new spiral phantom appears to meet the increased demands of spiral CT on phantom design, particularly z-axis length, and requirements for low contrast resolution test objects.

AB - Purpose: This paper reports on the development of a new phantom for spiral CT. The phantom meets the increased demands on phantom z-axis uniformity in order that objects from the CT slice, immediately above and belOW the CT slice of interest, do not contribute perturbing information to the reconstructed CT slice. Material and methods: The phantom depends on formulation of tissue-like materials that can be cast and produced in both geometric and anthropomorphic shapes with sufficient z-axis length to enable unperturbed CT slices of test objects of interest. These materials are then used to produce a series of test objects of CT image quality including low contrast samples that do not require volume averaging or mixing of solutions, and that can reflect sub-slice thickness test objects and supra-slice thickness test objects. Results: The overall phantom and its individual test objects provides meaningful tests of spiral CT image quality including slice sensitivity, CT number linearity and tests of high and low contrast resolution. Schematic designs and actual CT scans are shown. Conclusion: The new spiral phantom appears to meet the increased demands of spiral CT on phantom design, particularly z-axis length, and requirements for low contrast resolution test objects.

UR - http://www.scopus.com/inward/record.url?scp=0032075814&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0032075814&partnerID=8YFLogxK

U2 - 10.1016/S0895-6111(98)00020-2

DO - 10.1016/S0895-6111(98)00020-2

M3 - Article

C2 - 9740042

AN - SCOPUS:0032075814

VL - 22

SP - 247

EP - 255

JO - Computerized Medical Imaging and Graphics

JF - Computerized Medical Imaging and Graphics

SN - 0895-6111

IS - 3

ER -