Evaluation of Hemodynamic Performance of a Combined ECLS and CRRT Circuit in Seven Positions With a Simulated Neonatal Patient

Elizabeth Profeta, Kaitlyn Shank, Shigang Wang, Christian O'Connor, Allen Kunselman, Karl Woitas, John Myers, Akif Undar

Research output: Contribution to journalArticle

6 Citations (Scopus)

Abstract

As it is common for patients treated with extracorporeal life support (ECLS) to subsequently require continuous renal replacement therapy (CRRT), and neonatal patients encounter limitations due to lack of access points, inclusion of CRRT in the ECLS circuit could provide advanced treatment for this population. The objective of this study was to evaluate an alternative neonatal ECLS circuit containing either a Maquet RotaFlow centrifugal pump or Maquet HL20 roller pump with one of seven configurations of CRRT using the Prismaflex 2000 System. All ECLS circuit setups included a Quadrox-iD Pediatric diffusion membrane oxygenator, a Better Bladder, an 8-Fr arterial cannula, a 10-Fr venous cannula, and 6 feet of ¼-inch diameter arterial and venous tubing. The circuit was primed with lactated Ringer's solution and packed human red blood cells resulting in a total priming volume of 700 mL for both the circuit and the 3-kg pseudopatient. Hemodynamic data were recorded for ECLS flow rates of 200, 400, and 600 mL/min and a CRRT flow rate of 50 mL/min. When a centrifugal pump is used, the hemodynamic performance of any combined ECLS and CRRT circuit was not significantly different than that of the circuit without CRRT, thus any configuration could potentially be used. However, introduction of CRRT to a circuit containing a roller pump does affect performance properties for some CRRT positions. The circuits with CRRT positions B and G demonstrated decreased total hemodynamic energy (THE) levels at the post-arterial cannula site, while positions D and E demonstrated increased post-arterial cannula THE levels compared to the circuit without CRRT. CRRT positions A, C, and F did not have significant changes with respect to pre-arterial cannula flow and THE levels, compared to the circuit without CRRT. Considering hemodynamic performance, for neonatal combined extracorporeal membrane oxygenation (ECMO) and CRRT circuits with both blood pumps, we recommend the use of CRRT position A due to its hemodynamic similarities to the ECMO circuit without CRRT.

Original languageEnglish (US)
Pages (from-to)155-165
Number of pages11
JournalArtificial Organs
Volume42
Issue number2
DOIs
StatePublished - Feb 1 2018

Fingerprint

Extracorporeal Membrane Oxygenation
Renal Replacement Therapy
Hemodynamics
Networks (circuits)
Electron energy levels
Oxygenation
Centrifugal pumps
Pumps
Membranes
Blood
Flow rate
Oxygenators
Membrane Oxygenators
Pediatrics
Tubing

All Science Journal Classification (ASJC) codes

  • Bioengineering
  • Medicine (miscellaneous)
  • Biomaterials
  • Biomedical Engineering

Cite this

Profeta, Elizabeth ; Shank, Kaitlyn ; Wang, Shigang ; O'Connor, Christian ; Kunselman, Allen ; Woitas, Karl ; Myers, John ; Undar, Akif. / Evaluation of Hemodynamic Performance of a Combined ECLS and CRRT Circuit in Seven Positions With a Simulated Neonatal Patient. In: Artificial Organs. 2018 ; Vol. 42, No. 2. pp. 155-165.
@article{3f403b3ebf65430684bdb364470de1b5,
title = "Evaluation of Hemodynamic Performance of a Combined ECLS and CRRT Circuit in Seven Positions With a Simulated Neonatal Patient",
abstract = "As it is common for patients treated with extracorporeal life support (ECLS) to subsequently require continuous renal replacement therapy (CRRT), and neonatal patients encounter limitations due to lack of access points, inclusion of CRRT in the ECLS circuit could provide advanced treatment for this population. The objective of this study was to evaluate an alternative neonatal ECLS circuit containing either a Maquet RotaFlow centrifugal pump or Maquet HL20 roller pump with one of seven configurations of CRRT using the Prismaflex 2000 System. All ECLS circuit setups included a Quadrox-iD Pediatric diffusion membrane oxygenator, a Better Bladder, an 8-Fr arterial cannula, a 10-Fr venous cannula, and 6 feet of ¼-inch diameter arterial and venous tubing. The circuit was primed with lactated Ringer's solution and packed human red blood cells resulting in a total priming volume of 700 mL for both the circuit and the 3-kg pseudopatient. Hemodynamic data were recorded for ECLS flow rates of 200, 400, and 600 mL/min and a CRRT flow rate of 50 mL/min. When a centrifugal pump is used, the hemodynamic performance of any combined ECLS and CRRT circuit was not significantly different than that of the circuit without CRRT, thus any configuration could potentially be used. However, introduction of CRRT to a circuit containing a roller pump does affect performance properties for some CRRT positions. The circuits with CRRT positions B and G demonstrated decreased total hemodynamic energy (THE) levels at the post-arterial cannula site, while positions D and E demonstrated increased post-arterial cannula THE levels compared to the circuit without CRRT. CRRT positions A, C, and F did not have significant changes with respect to pre-arterial cannula flow and THE levels, compared to the circuit without CRRT. Considering hemodynamic performance, for neonatal combined extracorporeal membrane oxygenation (ECMO) and CRRT circuits with both blood pumps, we recommend the use of CRRT position A due to its hemodynamic similarities to the ECMO circuit without CRRT.",
author = "Elizabeth Profeta and Kaitlyn Shank and Shigang Wang and Christian O'Connor and Allen Kunselman and Karl Woitas and John Myers and Akif Undar",
year = "2018",
month = "2",
day = "1",
doi = "10.1111/aor.12907",
language = "English (US)",
volume = "42",
pages = "155--165",
journal = "Artificial Organs",
issn = "0160-564X",
publisher = "Wiley-Blackwell",
number = "2",

}

Evaluation of Hemodynamic Performance of a Combined ECLS and CRRT Circuit in Seven Positions With a Simulated Neonatal Patient. / Profeta, Elizabeth; Shank, Kaitlyn; Wang, Shigang; O'Connor, Christian; Kunselman, Allen; Woitas, Karl; Myers, John; Undar, Akif.

In: Artificial Organs, Vol. 42, No. 2, 01.02.2018, p. 155-165.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Evaluation of Hemodynamic Performance of a Combined ECLS and CRRT Circuit in Seven Positions With a Simulated Neonatal Patient

AU - Profeta, Elizabeth

AU - Shank, Kaitlyn

AU - Wang, Shigang

AU - O'Connor, Christian

AU - Kunselman, Allen

AU - Woitas, Karl

AU - Myers, John

AU - Undar, Akif

PY - 2018/2/1

Y1 - 2018/2/1

N2 - As it is common for patients treated with extracorporeal life support (ECLS) to subsequently require continuous renal replacement therapy (CRRT), and neonatal patients encounter limitations due to lack of access points, inclusion of CRRT in the ECLS circuit could provide advanced treatment for this population. The objective of this study was to evaluate an alternative neonatal ECLS circuit containing either a Maquet RotaFlow centrifugal pump or Maquet HL20 roller pump with one of seven configurations of CRRT using the Prismaflex 2000 System. All ECLS circuit setups included a Quadrox-iD Pediatric diffusion membrane oxygenator, a Better Bladder, an 8-Fr arterial cannula, a 10-Fr venous cannula, and 6 feet of ¼-inch diameter arterial and venous tubing. The circuit was primed with lactated Ringer's solution and packed human red blood cells resulting in a total priming volume of 700 mL for both the circuit and the 3-kg pseudopatient. Hemodynamic data were recorded for ECLS flow rates of 200, 400, and 600 mL/min and a CRRT flow rate of 50 mL/min. When a centrifugal pump is used, the hemodynamic performance of any combined ECLS and CRRT circuit was not significantly different than that of the circuit without CRRT, thus any configuration could potentially be used. However, introduction of CRRT to a circuit containing a roller pump does affect performance properties for some CRRT positions. The circuits with CRRT positions B and G demonstrated decreased total hemodynamic energy (THE) levels at the post-arterial cannula site, while positions D and E demonstrated increased post-arterial cannula THE levels compared to the circuit without CRRT. CRRT positions A, C, and F did not have significant changes with respect to pre-arterial cannula flow and THE levels, compared to the circuit without CRRT. Considering hemodynamic performance, for neonatal combined extracorporeal membrane oxygenation (ECMO) and CRRT circuits with both blood pumps, we recommend the use of CRRT position A due to its hemodynamic similarities to the ECMO circuit without CRRT.

AB - As it is common for patients treated with extracorporeal life support (ECLS) to subsequently require continuous renal replacement therapy (CRRT), and neonatal patients encounter limitations due to lack of access points, inclusion of CRRT in the ECLS circuit could provide advanced treatment for this population. The objective of this study was to evaluate an alternative neonatal ECLS circuit containing either a Maquet RotaFlow centrifugal pump or Maquet HL20 roller pump with one of seven configurations of CRRT using the Prismaflex 2000 System. All ECLS circuit setups included a Quadrox-iD Pediatric diffusion membrane oxygenator, a Better Bladder, an 8-Fr arterial cannula, a 10-Fr venous cannula, and 6 feet of ¼-inch diameter arterial and venous tubing. The circuit was primed with lactated Ringer's solution and packed human red blood cells resulting in a total priming volume of 700 mL for both the circuit and the 3-kg pseudopatient. Hemodynamic data were recorded for ECLS flow rates of 200, 400, and 600 mL/min and a CRRT flow rate of 50 mL/min. When a centrifugal pump is used, the hemodynamic performance of any combined ECLS and CRRT circuit was not significantly different than that of the circuit without CRRT, thus any configuration could potentially be used. However, introduction of CRRT to a circuit containing a roller pump does affect performance properties for some CRRT positions. The circuits with CRRT positions B and G demonstrated decreased total hemodynamic energy (THE) levels at the post-arterial cannula site, while positions D and E demonstrated increased post-arterial cannula THE levels compared to the circuit without CRRT. CRRT positions A, C, and F did not have significant changes with respect to pre-arterial cannula flow and THE levels, compared to the circuit without CRRT. Considering hemodynamic performance, for neonatal combined extracorporeal membrane oxygenation (ECMO) and CRRT circuits with both blood pumps, we recommend the use of CRRT position A due to its hemodynamic similarities to the ECMO circuit without CRRT.

UR - http://www.scopus.com/inward/record.url?scp=85021402426&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85021402426&partnerID=8YFLogxK

U2 - 10.1111/aor.12907

DO - 10.1111/aor.12907

M3 - Article

C2 - 28621839

AN - SCOPUS:85021402426

VL - 42

SP - 155

EP - 165

JO - Artificial Organs

JF - Artificial Organs

SN - 0160-564X

IS - 2

ER -