Flow characteristics of cerebrospinal fluid shunt tubing: Laboratory investigation

Joseph T. Cheatle, Alexis N. Bowder, Sandeep K. Agrawal, Michael Sather, Leslie C. Hellbusch

Research output: Contribution to journalArticle

6 Citations (Scopus)

Abstract

Object. Cerebrospinal fluid shunt systems malfunction for a multitude of reasons, including malpostitioning, obstruction of the ventricular or distal catheter, obstruction of the shunt valve, and catheter disruptions or disconnections. The goal of this study was to examine the hydrodynamic resistance and flow in new and explanted catheters and also in catheters with 1 or 2 straight connectors. Methods. Explanted catheters of multiple lengths, 2-piece catheters, 3-piece catheters, and new catheters were attached to a proximal and distal manometer. A flask with artificial CSF attached to the proximal end provided flow. The flow was allowed to stabilize over 1 hour; then the change in pressure between the proximal and distal end of the catheter was measured. Results. The resistance to flow was calculated for new, never-implanted catheters and compared with the resistance of explanted distal shunt catheters. The resistance of the new catheters was examined after the addition of 1 and 2 straight connectors. Explanted catheters exhibited a slight increase in the resistance to flow of artificial CSF compared with new catheters. Two-piece and 3-piece catheters had a significant increase in resistance to flow compared with new catheters. For all catheters, resistance to flow increased as length increased (new, p = 0.01; explanted, p = 0.009; 1 connector, p = 0.01; 2 connectors, p = 0.03). In this paper, effective diameter is defined as the available cross-sectional area of catheter contacted by the artificial CSF. For new and explanted catheters, a decrease in the effective diameter of the catheter was associated with an increase in the resistance to flow of artificial CSF (new, p = 0.1083; explanted, p = 0.0091). However, after the addition of 1 or 2 connectors, an inverse trend was observed: resistance to flow increased with effective diameter. Conclusions. There appears to be some increase in resistance of CSF shunt catheters as they age, altering flow dynamics. In addition, the use of straight connectors within a CSF shunt system increases the resistance to flow of artificial CSF within the shunt system. The increase in resistance appears to be related to the duration of implantation and the length of the catheter and inversely related to the diameter of the catheter. This increase in resistance may be related to sterile shunt malfunction. The addition of straight connectors is associated with a significant increase in resistance in comparison with catheters without connectors (p = 0.005).

Original languageEnglish (US)
Pages (from-to)191-197
Number of pages7
JournalJournal of Neurosurgery: Pediatrics
Volume9
Issue number2
DOIs
StatePublished - Feb 1 2012

Fingerprint

Cerebrospinal Fluid Shunts
Catheters

All Science Journal Classification (ASJC) codes

  • Clinical Neurology
  • Surgery
  • Pediatrics, Perinatology, and Child Health

Cite this

Cheatle, Joseph T. ; Bowder, Alexis N. ; Agrawal, Sandeep K. ; Sather, Michael ; Hellbusch, Leslie C. / Flow characteristics of cerebrospinal fluid shunt tubing : Laboratory investigation. In: Journal of Neurosurgery: Pediatrics. 2012 ; Vol. 9, No. 2. pp. 191-197.
@article{d562773154094a6cb187b4bb6d8a9657,
title = "Flow characteristics of cerebrospinal fluid shunt tubing: Laboratory investigation",
abstract = "Object. Cerebrospinal fluid shunt systems malfunction for a multitude of reasons, including malpostitioning, obstruction of the ventricular or distal catheter, obstruction of the shunt valve, and catheter disruptions or disconnections. The goal of this study was to examine the hydrodynamic resistance and flow in new and explanted catheters and also in catheters with 1 or 2 straight connectors. Methods. Explanted catheters of multiple lengths, 2-piece catheters, 3-piece catheters, and new catheters were attached to a proximal and distal manometer. A flask with artificial CSF attached to the proximal end provided flow. The flow was allowed to stabilize over 1 hour; then the change in pressure between the proximal and distal end of the catheter was measured. Results. The resistance to flow was calculated for new, never-implanted catheters and compared with the resistance of explanted distal shunt catheters. The resistance of the new catheters was examined after the addition of 1 and 2 straight connectors. Explanted catheters exhibited a slight increase in the resistance to flow of artificial CSF compared with new catheters. Two-piece and 3-piece catheters had a significant increase in resistance to flow compared with new catheters. For all catheters, resistance to flow increased as length increased (new, p = 0.01; explanted, p = 0.009; 1 connector, p = 0.01; 2 connectors, p = 0.03). In this paper, effective diameter is defined as the available cross-sectional area of catheter contacted by the artificial CSF. For new and explanted catheters, a decrease in the effective diameter of the catheter was associated with an increase in the resistance to flow of artificial CSF (new, p = 0.1083; explanted, p = 0.0091). However, after the addition of 1 or 2 connectors, an inverse trend was observed: resistance to flow increased with effective diameter. Conclusions. There appears to be some increase in resistance of CSF shunt catheters as they age, altering flow dynamics. In addition, the use of straight connectors within a CSF shunt system increases the resistance to flow of artificial CSF within the shunt system. The increase in resistance appears to be related to the duration of implantation and the length of the catheter and inversely related to the diameter of the catheter. This increase in resistance may be related to sterile shunt malfunction. The addition of straight connectors is associated with a significant increase in resistance in comparison with catheters without connectors (p = 0.005).",
author = "Cheatle, {Joseph T.} and Bowder, {Alexis N.} and Agrawal, {Sandeep K.} and Michael Sather and Hellbusch, {Leslie C.}",
year = "2012",
month = "2",
day = "1",
doi = "10.3171/2011.11.PEDS11255",
language = "English (US)",
volume = "9",
pages = "191--197",
journal = "Journal of Neurosurgery: Pediatrics",
issn = "1933-0707",
publisher = "American Association of Neurological Surgeons",
number = "2",

}

Flow characteristics of cerebrospinal fluid shunt tubing : Laboratory investigation. / Cheatle, Joseph T.; Bowder, Alexis N.; Agrawal, Sandeep K.; Sather, Michael; Hellbusch, Leslie C.

In: Journal of Neurosurgery: Pediatrics, Vol. 9, No. 2, 01.02.2012, p. 191-197.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Flow characteristics of cerebrospinal fluid shunt tubing

T2 - Laboratory investigation

AU - Cheatle, Joseph T.

AU - Bowder, Alexis N.

AU - Agrawal, Sandeep K.

AU - Sather, Michael

AU - Hellbusch, Leslie C.

PY - 2012/2/1

Y1 - 2012/2/1

N2 - Object. Cerebrospinal fluid shunt systems malfunction for a multitude of reasons, including malpostitioning, obstruction of the ventricular or distal catheter, obstruction of the shunt valve, and catheter disruptions or disconnections. The goal of this study was to examine the hydrodynamic resistance and flow in new and explanted catheters and also in catheters with 1 or 2 straight connectors. Methods. Explanted catheters of multiple lengths, 2-piece catheters, 3-piece catheters, and new catheters were attached to a proximal and distal manometer. A flask with artificial CSF attached to the proximal end provided flow. The flow was allowed to stabilize over 1 hour; then the change in pressure between the proximal and distal end of the catheter was measured. Results. The resistance to flow was calculated for new, never-implanted catheters and compared with the resistance of explanted distal shunt catheters. The resistance of the new catheters was examined after the addition of 1 and 2 straight connectors. Explanted catheters exhibited a slight increase in the resistance to flow of artificial CSF compared with new catheters. Two-piece and 3-piece catheters had a significant increase in resistance to flow compared with new catheters. For all catheters, resistance to flow increased as length increased (new, p = 0.01; explanted, p = 0.009; 1 connector, p = 0.01; 2 connectors, p = 0.03). In this paper, effective diameter is defined as the available cross-sectional area of catheter contacted by the artificial CSF. For new and explanted catheters, a decrease in the effective diameter of the catheter was associated with an increase in the resistance to flow of artificial CSF (new, p = 0.1083; explanted, p = 0.0091). However, after the addition of 1 or 2 connectors, an inverse trend was observed: resistance to flow increased with effective diameter. Conclusions. There appears to be some increase in resistance of CSF shunt catheters as they age, altering flow dynamics. In addition, the use of straight connectors within a CSF shunt system increases the resistance to flow of artificial CSF within the shunt system. The increase in resistance appears to be related to the duration of implantation and the length of the catheter and inversely related to the diameter of the catheter. This increase in resistance may be related to sterile shunt malfunction. The addition of straight connectors is associated with a significant increase in resistance in comparison with catheters without connectors (p = 0.005).

AB - Object. Cerebrospinal fluid shunt systems malfunction for a multitude of reasons, including malpostitioning, obstruction of the ventricular or distal catheter, obstruction of the shunt valve, and catheter disruptions or disconnections. The goal of this study was to examine the hydrodynamic resistance and flow in new and explanted catheters and also in catheters with 1 or 2 straight connectors. Methods. Explanted catheters of multiple lengths, 2-piece catheters, 3-piece catheters, and new catheters were attached to a proximal and distal manometer. A flask with artificial CSF attached to the proximal end provided flow. The flow was allowed to stabilize over 1 hour; then the change in pressure between the proximal and distal end of the catheter was measured. Results. The resistance to flow was calculated for new, never-implanted catheters and compared with the resistance of explanted distal shunt catheters. The resistance of the new catheters was examined after the addition of 1 and 2 straight connectors. Explanted catheters exhibited a slight increase in the resistance to flow of artificial CSF compared with new catheters. Two-piece and 3-piece catheters had a significant increase in resistance to flow compared with new catheters. For all catheters, resistance to flow increased as length increased (new, p = 0.01; explanted, p = 0.009; 1 connector, p = 0.01; 2 connectors, p = 0.03). In this paper, effective diameter is defined as the available cross-sectional area of catheter contacted by the artificial CSF. For new and explanted catheters, a decrease in the effective diameter of the catheter was associated with an increase in the resistance to flow of artificial CSF (new, p = 0.1083; explanted, p = 0.0091). However, after the addition of 1 or 2 connectors, an inverse trend was observed: resistance to flow increased with effective diameter. Conclusions. There appears to be some increase in resistance of CSF shunt catheters as they age, altering flow dynamics. In addition, the use of straight connectors within a CSF shunt system increases the resistance to flow of artificial CSF within the shunt system. The increase in resistance appears to be related to the duration of implantation and the length of the catheter and inversely related to the diameter of the catheter. This increase in resistance may be related to sterile shunt malfunction. The addition of straight connectors is associated with a significant increase in resistance in comparison with catheters without connectors (p = 0.005).

UR - http://www.scopus.com/inward/record.url?scp=84856795880&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84856795880&partnerID=8YFLogxK

U2 - 10.3171/2011.11.PEDS11255

DO - 10.3171/2011.11.PEDS11255

M3 - Article

C2 - 22295926

AN - SCOPUS:84856795880

VL - 9

SP - 191

EP - 197

JO - Journal of Neurosurgery: Pediatrics

JF - Journal of Neurosurgery: Pediatrics

SN - 1933-0707

IS - 2

ER -