Intraaortic balloon pump timing discrepancies in adult patients

George M. Pantalos, Steven C. Koenig, Kevin J. Gillars, Gilbert S. Haugh, Robert Dowling, Laman A. Gray

Research output: Contribution to journalArticle

7 Citations (Scopus)

Abstract

The objective of this clinical study was to quantify the incidence and magnitude of intraaortic balloon pump (IABP) inflation and deflation landmark discrepancies associated with the IABP catheter arterial pressure waveform. Cardiac surgery patients with an IABP inserted prior to surgery were recruited. Following cardiac exposure, a high-fidelity pressure catheter was inserted into the aortic root for digital recording. The radial artery pressure signal was simultaneously recorded from the patient monitor along with the arterial pressure and electrocardiogram waveforms from the IABP console while operating at 1:1 and 1:2 synchronization. In selected patients, recordings were obtained with the IABP timed to the high-fidelity aortic root waveform. In all 11 patients, inflation and deflation landmark delays were observed when comparing the aortic root waveforms to the IABP arterial pressure waveforms (inflation delay=74±29 [23-117] ms; deflation delay=71±37 [24-141] ms, mean±standard deviation [min-max]). Delays were greater when compared to the radial artery waveform (inflation delay=175±50 [100-233] ms; deflation delay=168±52 [100-274] ms). In all cases, the landmark delays were statistically different from zero (P<0.001). Diastolic augmentation and afterload reduction varied with waveform source. Conflicting indications of afterload reduction occurred in four patients. Timing to the aortic root waveform resulted in greater diastolic pressure augmentation and afterload reduction but mixed changes in stroke volume. Delay and distortion of the arterial waveform was consistently found when measured through the IABP catheter lumen. These delays can alter IABP efficacy and may be eliminated by using high-fidelity sensing of aortic pressure.

Original languageEnglish (US)
Pages (from-to)857-866
Number of pages10
JournalArtificial Organs
Volume35
Issue number9
DOIs
StatePublished - Sep 1 2011

Fingerprint

Balloons
Economic Inflation
Pumps
Arterial Pressure
Radial Artery
Catheters
Pressure
Surgery
Stroke Volume
Thoracic Surgery
Electrocardiography
Blood Pressure
Incidence
Synchronization

All Science Journal Classification (ASJC) codes

  • Bioengineering
  • Medicine (miscellaneous)
  • Biomaterials
  • Biomedical Engineering

Cite this

Pantalos, G. M., Koenig, S. C., Gillars, K. J., Haugh, G. S., Dowling, R., & Gray, L. A. (2011). Intraaortic balloon pump timing discrepancies in adult patients. Artificial Organs, 35(9), 857-866. https://doi.org/10.1111/j.1525-1594.2011.01292.x
Pantalos, George M. ; Koenig, Steven C. ; Gillars, Kevin J. ; Haugh, Gilbert S. ; Dowling, Robert ; Gray, Laman A. / Intraaortic balloon pump timing discrepancies in adult patients. In: Artificial Organs. 2011 ; Vol. 35, No. 9. pp. 857-866.
@article{e1c1ffbeef4b49bda07efb7e6f24c144,
title = "Intraaortic balloon pump timing discrepancies in adult patients",
abstract = "The objective of this clinical study was to quantify the incidence and magnitude of intraaortic balloon pump (IABP) inflation and deflation landmark discrepancies associated with the IABP catheter arterial pressure waveform. Cardiac surgery patients with an IABP inserted prior to surgery were recruited. Following cardiac exposure, a high-fidelity pressure catheter was inserted into the aortic root for digital recording. The radial artery pressure signal was simultaneously recorded from the patient monitor along with the arterial pressure and electrocardiogram waveforms from the IABP console while operating at 1:1 and 1:2 synchronization. In selected patients, recordings were obtained with the IABP timed to the high-fidelity aortic root waveform. In all 11 patients, inflation and deflation landmark delays were observed when comparing the aortic root waveforms to the IABP arterial pressure waveforms (inflation delay=74±29 [23-117] ms; deflation delay=71±37 [24-141] ms, mean±standard deviation [min-max]). Delays were greater when compared to the radial artery waveform (inflation delay=175±50 [100-233] ms; deflation delay=168±52 [100-274] ms). In all cases, the landmark delays were statistically different from zero (P<0.001). Diastolic augmentation and afterload reduction varied with waveform source. Conflicting indications of afterload reduction occurred in four patients. Timing to the aortic root waveform resulted in greater diastolic pressure augmentation and afterload reduction but mixed changes in stroke volume. Delay and distortion of the arterial waveform was consistently found when measured through the IABP catheter lumen. These delays can alter IABP efficacy and may be eliminated by using high-fidelity sensing of aortic pressure.",
author = "Pantalos, {George M.} and Koenig, {Steven C.} and Gillars, {Kevin J.} and Haugh, {Gilbert S.} and Robert Dowling and Gray, {Laman A.}",
year = "2011",
month = "9",
day = "1",
doi = "10.1111/j.1525-1594.2011.01292.x",
language = "English (US)",
volume = "35",
pages = "857--866",
journal = "Artificial Organs",
issn = "0160-564X",
publisher = "Wiley-Blackwell",
number = "9",

}

Pantalos, GM, Koenig, SC, Gillars, KJ, Haugh, GS, Dowling, R & Gray, LA 2011, 'Intraaortic balloon pump timing discrepancies in adult patients', Artificial Organs, vol. 35, no. 9, pp. 857-866. https://doi.org/10.1111/j.1525-1594.2011.01292.x

Intraaortic balloon pump timing discrepancies in adult patients. / Pantalos, George M.; Koenig, Steven C.; Gillars, Kevin J.; Haugh, Gilbert S.; Dowling, Robert; Gray, Laman A.

In: Artificial Organs, Vol. 35, No. 9, 01.09.2011, p. 857-866.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Intraaortic balloon pump timing discrepancies in adult patients

AU - Pantalos, George M.

AU - Koenig, Steven C.

AU - Gillars, Kevin J.

AU - Haugh, Gilbert S.

AU - Dowling, Robert

AU - Gray, Laman A.

PY - 2011/9/1

Y1 - 2011/9/1

N2 - The objective of this clinical study was to quantify the incidence and magnitude of intraaortic balloon pump (IABP) inflation and deflation landmark discrepancies associated with the IABP catheter arterial pressure waveform. Cardiac surgery patients with an IABP inserted prior to surgery were recruited. Following cardiac exposure, a high-fidelity pressure catheter was inserted into the aortic root for digital recording. The radial artery pressure signal was simultaneously recorded from the patient monitor along with the arterial pressure and electrocardiogram waveforms from the IABP console while operating at 1:1 and 1:2 synchronization. In selected patients, recordings were obtained with the IABP timed to the high-fidelity aortic root waveform. In all 11 patients, inflation and deflation landmark delays were observed when comparing the aortic root waveforms to the IABP arterial pressure waveforms (inflation delay=74±29 [23-117] ms; deflation delay=71±37 [24-141] ms, mean±standard deviation [min-max]). Delays were greater when compared to the radial artery waveform (inflation delay=175±50 [100-233] ms; deflation delay=168±52 [100-274] ms). In all cases, the landmark delays were statistically different from zero (P<0.001). Diastolic augmentation and afterload reduction varied with waveform source. Conflicting indications of afterload reduction occurred in four patients. Timing to the aortic root waveform resulted in greater diastolic pressure augmentation and afterload reduction but mixed changes in stroke volume. Delay and distortion of the arterial waveform was consistently found when measured through the IABP catheter lumen. These delays can alter IABP efficacy and may be eliminated by using high-fidelity sensing of aortic pressure.

AB - The objective of this clinical study was to quantify the incidence and magnitude of intraaortic balloon pump (IABP) inflation and deflation landmark discrepancies associated with the IABP catheter arterial pressure waveform. Cardiac surgery patients with an IABP inserted prior to surgery were recruited. Following cardiac exposure, a high-fidelity pressure catheter was inserted into the aortic root for digital recording. The radial artery pressure signal was simultaneously recorded from the patient monitor along with the arterial pressure and electrocardiogram waveforms from the IABP console while operating at 1:1 and 1:2 synchronization. In selected patients, recordings were obtained with the IABP timed to the high-fidelity aortic root waveform. In all 11 patients, inflation and deflation landmark delays were observed when comparing the aortic root waveforms to the IABP arterial pressure waveforms (inflation delay=74±29 [23-117] ms; deflation delay=71±37 [24-141] ms, mean±standard deviation [min-max]). Delays were greater when compared to the radial artery waveform (inflation delay=175±50 [100-233] ms; deflation delay=168±52 [100-274] ms). In all cases, the landmark delays were statistically different from zero (P<0.001). Diastolic augmentation and afterload reduction varied with waveform source. Conflicting indications of afterload reduction occurred in four patients. Timing to the aortic root waveform resulted in greater diastolic pressure augmentation and afterload reduction but mixed changes in stroke volume. Delay and distortion of the arterial waveform was consistently found when measured through the IABP catheter lumen. These delays can alter IABP efficacy and may be eliminated by using high-fidelity sensing of aortic pressure.

UR - http://www.scopus.com/inward/record.url?scp=80052681267&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=80052681267&partnerID=8YFLogxK

U2 - 10.1111/j.1525-1594.2011.01292.x

DO - 10.1111/j.1525-1594.2011.01292.x

M3 - Article

C2 - 21883317

AN - SCOPUS:80052681267

VL - 35

SP - 857

EP - 866

JO - Artificial Organs

JF - Artificial Organs

SN - 0160-564X

IS - 9

ER -