Abstract

In this paper, a state-space model is developed through theoretical analysis and numerical solutions to approximate the response of the human circulatory system. This system model has one critical time-varying parameter: the resistance of peripheral blood vessels. A parameter estimation scheme is derived to estimate this parameter, and the parameter estimate is used to implement an adaptive observer to estimate the aortic pressure for physiological control. An optimal adaptive controller is proposed to control the estimated aortic pressure to track a reference signal updated by a nonlinear function of the pump head to meet the physiological need. A Matlab simulation program and an experimental mock human circulatory loop are employed as test environments for the human circulatory systems with a left ventricular assist device and their physiological controllers. Different physiological conditions, such as the variation of left ventricular failures, variation of activities, and collapse of the left ventricle, are evaluated to test the designed physiological control system. Simulation and experimental results consistently show that the aortic pressure estimation error is small, and that the abnormal hemodynamic variables of a congestive heart failure patient are restored back to the normal physiological range.

Original languageEnglish (US)
Pages (from-to)754-767
Number of pages14
JournalIEEE Transactions on Control Systems Technology
Volume15
Issue number4
DOIs
StatePublished - Jul 1 2007

Fingerprint

Left ventricular assist devices
Controllers
Blood vessels
Hemodynamics
Parameter estimation
Error analysis
Pumps
Control systems

All Science Journal Classification (ASJC) codes

  • Control and Systems Engineering
  • Electrical and Electronic Engineering

Cite this

@article{b927f5e197134848aef614c545ad31ff,
title = "Modeling, estimation, and control of human circulatory system with a left ventricular assist device",
abstract = "In this paper, a state-space model is developed through theoretical analysis and numerical solutions to approximate the response of the human circulatory system. This system model has one critical time-varying parameter: the resistance of peripheral blood vessels. A parameter estimation scheme is derived to estimate this parameter, and the parameter estimate is used to implement an adaptive observer to estimate the aortic pressure for physiological control. An optimal adaptive controller is proposed to control the estimated aortic pressure to track a reference signal updated by a nonlinear function of the pump head to meet the physiological need. A Matlab simulation program and an experimental mock human circulatory loop are employed as test environments for the human circulatory systems with a left ventricular assist device and their physiological controllers. Different physiological conditions, such as the variation of left ventricular failures, variation of activities, and collapse of the left ventricle, are evaluated to test the designed physiological control system. Simulation and experimental results consistently show that the aortic pressure estimation error is small, and that the abnormal hemodynamic variables of a congestive heart failure patient are restored back to the normal physiological range.",
author = "Yi Wu and Allaire, {Paul E.} and Gang Tao and Don Olsen",
year = "2007",
month = "7",
day = "1",
doi = "10.1109/TCST.2006.890288",
language = "English (US)",
volume = "15",
pages = "754--767",
journal = "IEEE Transactions on Control Systems Technology",
issn = "1063-6536",
publisher = "Institute of Electrical and Electronics Engineers Inc.",
number = "4",

}

Modeling, estimation, and control of human circulatory system with a left ventricular assist device. / Wu, Yi; Allaire, Paul E.; Tao, Gang; Olsen, Don.

In: IEEE Transactions on Control Systems Technology, Vol. 15, No. 4, 01.07.2007, p. 754-767.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Modeling, estimation, and control of human circulatory system with a left ventricular assist device

AU - Wu, Yi

AU - Allaire, Paul E.

AU - Tao, Gang

AU - Olsen, Don

PY - 2007/7/1

Y1 - 2007/7/1

N2 - In this paper, a state-space model is developed through theoretical analysis and numerical solutions to approximate the response of the human circulatory system. This system model has one critical time-varying parameter: the resistance of peripheral blood vessels. A parameter estimation scheme is derived to estimate this parameter, and the parameter estimate is used to implement an adaptive observer to estimate the aortic pressure for physiological control. An optimal adaptive controller is proposed to control the estimated aortic pressure to track a reference signal updated by a nonlinear function of the pump head to meet the physiological need. A Matlab simulation program and an experimental mock human circulatory loop are employed as test environments for the human circulatory systems with a left ventricular assist device and their physiological controllers. Different physiological conditions, such as the variation of left ventricular failures, variation of activities, and collapse of the left ventricle, are evaluated to test the designed physiological control system. Simulation and experimental results consistently show that the aortic pressure estimation error is small, and that the abnormal hemodynamic variables of a congestive heart failure patient are restored back to the normal physiological range.

AB - In this paper, a state-space model is developed through theoretical analysis and numerical solutions to approximate the response of the human circulatory system. This system model has one critical time-varying parameter: the resistance of peripheral blood vessels. A parameter estimation scheme is derived to estimate this parameter, and the parameter estimate is used to implement an adaptive observer to estimate the aortic pressure for physiological control. An optimal adaptive controller is proposed to control the estimated aortic pressure to track a reference signal updated by a nonlinear function of the pump head to meet the physiological need. A Matlab simulation program and an experimental mock human circulatory loop are employed as test environments for the human circulatory systems with a left ventricular assist device and their physiological controllers. Different physiological conditions, such as the variation of left ventricular failures, variation of activities, and collapse of the left ventricle, are evaluated to test the designed physiological control system. Simulation and experimental results consistently show that the aortic pressure estimation error is small, and that the abnormal hemodynamic variables of a congestive heart failure patient are restored back to the normal physiological range.

UR - http://www.scopus.com/inward/record.url?scp=34447335841&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=34447335841&partnerID=8YFLogxK

U2 - 10.1109/TCST.2006.890288

DO - 10.1109/TCST.2006.890288

M3 - Article

AN - SCOPUS:34447335841

VL - 15

SP - 754

EP - 767

JO - IEEE Transactions on Control Systems Technology

JF - IEEE Transactions on Control Systems Technology

SN - 1063-6536

IS - 4

ER -