    The purpose of this study was to investigate the accuracy of ultrasound speckle tracking in various tissues. Results from two-dimensional tissue speckle tracking in liver, muscle, fat, and sponge samples are presented, while keeping other speckle tracking parameters constant. Speckle tracking performance was characterized both in terms of the magnitude of tracking errors and in terms of the percentage of correctly tracked displacement vectors. Speckle tracking in muscle tissue, which contains myofibrils and significant tissue microstructure, produced the highest percentage of correctly tracked vectors and smallest tracking errors relative to other tissues.

    Original languageEnglish (US)
    Pages (from-to)1273-1278
    Number of pages6
    JournalJournal of the Acoustical Society of America
    Volume98
    Issue number3
    DOIs
    StatePublished - Jan 1 1995

    Fingerprint

    ultrasonics
    muscles
    fats
    liver
    microstructure

    All Science Journal Classification (ASJC) codes

    • Arts and Humanities (miscellaneous)
    • Acoustics and Ultrasonics

    Cite this

    Chen, Eric J. ; Jenkins, William Kenneth ; O'Brien, William D. / Performance of ultrasonic speckle tracking in various tissuesa. In: Journal of the Acoustical Society of America. 1995 ; Vol. 98, No. 3. pp. 1273-1278.
    @article{06ead963fcbf4536afaec92d46cf36ac,
    title = "Performance of ultrasonic speckle tracking in various tissuesa",
    abstract = "The purpose of this study was to investigate the accuracy of ultrasound speckle tracking in various tissues. Results from two-dimensional tissue speckle tracking in liver, muscle, fat, and sponge samples are presented, while keeping other speckle tracking parameters constant. Speckle tracking performance was characterized both in terms of the magnitude of tracking errors and in terms of the percentage of correctly tracked displacement vectors. Speckle tracking in muscle tissue, which contains myofibrils and significant tissue microstructure, produced the highest percentage of correctly tracked vectors and smallest tracking errors relative to other tissues.",
    author = "Chen, {Eric J.} and Jenkins, {William Kenneth} and O'Brien, {William D.}",
    year = "1995",
    month = "1",
    day = "1",
    doi = "10.1121/1.414453",
    language = "English (US)",
    volume = "98",
    pages = "1273--1278",
    journal = "Journal of the Acoustical Society of America",
    issn = "0001-4966",
    publisher = "Acoustical Society of America",
    number = "3",

    }

    Performance of ultrasonic speckle tracking in various tissuesa. / Chen, Eric J.; Jenkins, William Kenneth; O'Brien, William D.

    In: Journal of the Acoustical Society of America, Vol. 98, No. 3, 01.01.1995, p. 1273-1278.

    Research output: Contribution to journalArticle

    TY - JOUR

    T1 - Performance of ultrasonic speckle tracking in various tissuesa

    AU - Chen, Eric J.

    AU - Jenkins, William Kenneth

    AU - O'Brien, William D.

    PY - 1995/1/1

    Y1 - 1995/1/1

    N2 - The purpose of this study was to investigate the accuracy of ultrasound speckle tracking in various tissues. Results from two-dimensional tissue speckle tracking in liver, muscle, fat, and sponge samples are presented, while keeping other speckle tracking parameters constant. Speckle tracking performance was characterized both in terms of the magnitude of tracking errors and in terms of the percentage of correctly tracked displacement vectors. Speckle tracking in muscle tissue, which contains myofibrils and significant tissue microstructure, produced the highest percentage of correctly tracked vectors and smallest tracking errors relative to other tissues.

    AB - The purpose of this study was to investigate the accuracy of ultrasound speckle tracking in various tissues. Results from two-dimensional tissue speckle tracking in liver, muscle, fat, and sponge samples are presented, while keeping other speckle tracking parameters constant. Speckle tracking performance was characterized both in terms of the magnitude of tracking errors and in terms of the percentage of correctly tracked displacement vectors. Speckle tracking in muscle tissue, which contains myofibrils and significant tissue microstructure, produced the highest percentage of correctly tracked vectors and smallest tracking errors relative to other tissues.

    UR - http://www.scopus.com/inward/record.url?scp=0029133425&partnerID=8YFLogxK

    UR - http://www.scopus.com/inward/citedby.url?scp=0029133425&partnerID=8YFLogxK

    U2 - 10.1121/1.414453

    DO - 10.1121/1.414453

    M3 - Article

    C2 - 7560501

    AN - SCOPUS:0029133425

    VL - 98

    SP - 1273

    EP - 1278

    JO - Journal of the Acoustical Society of America

    JF - Journal of the Acoustical Society of America

    SN - 0001-4966

    IS - 3

    ER -