The esophageal-tracheal combitube resistance and ventilatory pressures

Luis A. Gaitini, Sonia J. Vaida, Somri Mostafa, Boris Yanovski, Bruce Ben-David, Jonathan L. Benumof

Research output: Contribution to journalArticle

8 Citations (Scopus)

Abstract

Study Objective: To measure resistance of the Combitube, a supraglottic ventilatory device used in the management of the patients with difficult airways, and its influence on delivered ventilatory pressures. Design: Prospective study. Setting: University-affiliated hospital. Patients: A total of 20 patients with ASA status I or II who were scheduled for elective knee arthroscopy. Interventions: (Part 2 of the study) After induction of general anesthesia and insertion of the Combitube, mechanical ventilation was initiated. Airway pressures were measured using fluid-filled pressure lines at the Y-piece (PY-piece) of the breathing system and in the oropharynx (P oropharynx) at a position 2 cm beyond the second proximal anterior hole of the Combitube. These pressures were simultaneously recorded and the pressure curves were compared. Measurements: (Part 1 of the study) Resistance of the esophageal and the tracheal lumen of the 37-F Combitube and standard endotracheal tubes (with internal diameters of 6, 7, and 8 mm) was compared ex vivo with a Datex AS/3 monitor. Ventilation conditions were kept constant at a tidal volume of 0.5 L, frequency of 10 breaths per minute, and ramp flow waveform and peak flow of 1 L/s. Main Results: Resistance of standard endotracheal tubes was inversely proportional to their diameters (16, 11, and 7 cm H2O/L per second for the tubes with internal diameters of 6, 7, and 8 mm, respectively). The resistance of the Combitube's tracheal lumen was 12 cm H2O/L per second. There was a significant difference in peak respiratory pressures between PY-piece and Poropharynx (40 ± 5 and 23 ± 5 cm H2O, respectively). Conclusions: The Combitube has significant airflow resistance that should be considered when patients are mechanically ventilated because the delivered oropharyngeal pressure is significantly lower than the pressure measured at the anesthesia breathing system.

Original languageEnglish (US)
Pages (from-to)26-29
Number of pages4
JournalJournal of Clinical Anesthesia
Volume17
Issue number1
DOIs
StatePublished - Feb 2005

Fingerprint

Pressure
Oropharynx
Respiration
Architectural Accessibility
Arthroscopy
Tidal Volume
Artificial Respiration
General Anesthesia
Ventilation
Knee
Anesthesia
Prospective Studies
Equipment and Supplies

All Science Journal Classification (ASJC) codes

  • Anesthesiology and Pain Medicine

Cite this

Gaitini, Luis A. ; Vaida, Sonia J. ; Mostafa, Somri ; Yanovski, Boris ; Ben-David, Bruce ; Benumof, Jonathan L. / The esophageal-tracheal combitube resistance and ventilatory pressures. In: Journal of Clinical Anesthesia. 2005 ; Vol. 17, No. 1. pp. 26-29.
@article{83a1ab3bf9f94656b09a5a6db570943d,
title = "The esophageal-tracheal combitube resistance and ventilatory pressures",
abstract = "Study Objective: To measure resistance of the Combitube, a supraglottic ventilatory device used in the management of the patients with difficult airways, and its influence on delivered ventilatory pressures. Design: Prospective study. Setting: University-affiliated hospital. Patients: A total of 20 patients with ASA status I or II who were scheduled for elective knee arthroscopy. Interventions: (Part 2 of the study) After induction of general anesthesia and insertion of the Combitube, mechanical ventilation was initiated. Airway pressures were measured using fluid-filled pressure lines at the Y-piece (PY-piece) of the breathing system and in the oropharynx (P oropharynx) at a position 2 cm beyond the second proximal anterior hole of the Combitube. These pressures were simultaneously recorded and the pressure curves were compared. Measurements: (Part 1 of the study) Resistance of the esophageal and the tracheal lumen of the 37-F Combitube and standard endotracheal tubes (with internal diameters of 6, 7, and 8 mm) was compared ex vivo with a Datex AS/3 monitor. Ventilation conditions were kept constant at a tidal volume of 0.5 L, frequency of 10 breaths per minute, and ramp flow waveform and peak flow of 1 L/s. Main Results: Resistance of standard endotracheal tubes was inversely proportional to their diameters (16, 11, and 7 cm H2O/L per second for the tubes with internal diameters of 6, 7, and 8 mm, respectively). The resistance of the Combitube's tracheal lumen was 12 cm H2O/L per second. There was a significant difference in peak respiratory pressures between PY-piece and Poropharynx (40 ± 5 and 23 ± 5 cm H2O, respectively). Conclusions: The Combitube has significant airflow resistance that should be considered when patients are mechanically ventilated because the delivered oropharyngeal pressure is significantly lower than the pressure measured at the anesthesia breathing system.",
author = "Gaitini, {Luis A.} and Vaida, {Sonia J.} and Somri Mostafa and Boris Yanovski and Bruce Ben-David and Benumof, {Jonathan L.}",
year = "2005",
month = "2",
doi = "10.1016/j.jclinane.2004.03.010",
language = "English (US)",
volume = "17",
pages = "26--29",
journal = "Journal of Clinical Anesthesia",
issn = "0952-8180",
publisher = "Elsevier Inc.",
number = "1",

}

The esophageal-tracheal combitube resistance and ventilatory pressures. / Gaitini, Luis A.; Vaida, Sonia J.; Mostafa, Somri; Yanovski, Boris; Ben-David, Bruce; Benumof, Jonathan L.

In: Journal of Clinical Anesthesia, Vol. 17, No. 1, 02.2005, p. 26-29.

Research output: Contribution to journalArticle

TY - JOUR

T1 - The esophageal-tracheal combitube resistance and ventilatory pressures

AU - Gaitini, Luis A.

AU - Vaida, Sonia J.

AU - Mostafa, Somri

AU - Yanovski, Boris

AU - Ben-David, Bruce

AU - Benumof, Jonathan L.

PY - 2005/2

Y1 - 2005/2

N2 - Study Objective: To measure resistance of the Combitube, a supraglottic ventilatory device used in the management of the patients with difficult airways, and its influence on delivered ventilatory pressures. Design: Prospective study. Setting: University-affiliated hospital. Patients: A total of 20 patients with ASA status I or II who were scheduled for elective knee arthroscopy. Interventions: (Part 2 of the study) After induction of general anesthesia and insertion of the Combitube, mechanical ventilation was initiated. Airway pressures were measured using fluid-filled pressure lines at the Y-piece (PY-piece) of the breathing system and in the oropharynx (P oropharynx) at a position 2 cm beyond the second proximal anterior hole of the Combitube. These pressures were simultaneously recorded and the pressure curves were compared. Measurements: (Part 1 of the study) Resistance of the esophageal and the tracheal lumen of the 37-F Combitube and standard endotracheal tubes (with internal diameters of 6, 7, and 8 mm) was compared ex vivo with a Datex AS/3 monitor. Ventilation conditions were kept constant at a tidal volume of 0.5 L, frequency of 10 breaths per minute, and ramp flow waveform and peak flow of 1 L/s. Main Results: Resistance of standard endotracheal tubes was inversely proportional to their diameters (16, 11, and 7 cm H2O/L per second for the tubes with internal diameters of 6, 7, and 8 mm, respectively). The resistance of the Combitube's tracheal lumen was 12 cm H2O/L per second. There was a significant difference in peak respiratory pressures between PY-piece and Poropharynx (40 ± 5 and 23 ± 5 cm H2O, respectively). Conclusions: The Combitube has significant airflow resistance that should be considered when patients are mechanically ventilated because the delivered oropharyngeal pressure is significantly lower than the pressure measured at the anesthesia breathing system.

AB - Study Objective: To measure resistance of the Combitube, a supraglottic ventilatory device used in the management of the patients with difficult airways, and its influence on delivered ventilatory pressures. Design: Prospective study. Setting: University-affiliated hospital. Patients: A total of 20 patients with ASA status I or II who were scheduled for elective knee arthroscopy. Interventions: (Part 2 of the study) After induction of general anesthesia and insertion of the Combitube, mechanical ventilation was initiated. Airway pressures were measured using fluid-filled pressure lines at the Y-piece (PY-piece) of the breathing system and in the oropharynx (P oropharynx) at a position 2 cm beyond the second proximal anterior hole of the Combitube. These pressures were simultaneously recorded and the pressure curves were compared. Measurements: (Part 1 of the study) Resistance of the esophageal and the tracheal lumen of the 37-F Combitube and standard endotracheal tubes (with internal diameters of 6, 7, and 8 mm) was compared ex vivo with a Datex AS/3 monitor. Ventilation conditions were kept constant at a tidal volume of 0.5 L, frequency of 10 breaths per minute, and ramp flow waveform and peak flow of 1 L/s. Main Results: Resistance of standard endotracheal tubes was inversely proportional to their diameters (16, 11, and 7 cm H2O/L per second for the tubes with internal diameters of 6, 7, and 8 mm, respectively). The resistance of the Combitube's tracheal lumen was 12 cm H2O/L per second. There was a significant difference in peak respiratory pressures between PY-piece and Poropharynx (40 ± 5 and 23 ± 5 cm H2O, respectively). Conclusions: The Combitube has significant airflow resistance that should be considered when patients are mechanically ventilated because the delivered oropharyngeal pressure is significantly lower than the pressure measured at the anesthesia breathing system.

UR - http://www.scopus.com/inward/record.url?scp=13844255235&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=13844255235&partnerID=8YFLogxK

U2 - 10.1016/j.jclinane.2004.03.010

DO - 10.1016/j.jclinane.2004.03.010

M3 - Article

C2 - 15721726

AN - SCOPUS:13844255235

VL - 17

SP - 26

EP - 29

JO - Journal of Clinical Anesthesia

JF - Journal of Clinical Anesthesia

SN - 0952-8180

IS - 1