Abstract

Experimental temperature profiles and biological kill data from dead-ended tubes of various diameters and lengths commonly used in biotechnology applications were compared to those predicted by a finite element model for steam-in-place (SIP) sterilization at 122°C. Diameter was shown experimentally and numerically to have a significant effect with larger diameter tubes exhibiting greater buoyant driven convective flow and more rapid sterilization. The overall Grashof number was shown to be the significant parameter relating magnitude of convective flow to tube diameter and varied as the diameter cubed. Analysis of air/steam mixture flow patterns showed air displacement from 0.4 cm diameter tubes to be due primarily to molecular diffusion, whereas 1.0 and 1.7 cm tubes showed a two-stage convective flow pattern. There exists a critical diameter of 0.4 cm below which SIP sterilization due to buoyancy driven flow does not occur and steam bleeders should be used.

Original languageEnglish (US)
Pages (from-to)293-304
Number of pages12
JournalBioprocess Engineering
Volume12
Issue number6
DOIs
StatePublished - May 1 1995

Fingerprint

Steam
Leg
Flow patterns
Air
Grashof number
Biotechnology
Buoyancy
Temperature

All Science Journal Classification (ASJC) codes

  • Biotechnology
  • Applied Microbiology and Biotechnology

Cite this

Young, J. H. ; Lasher, William ; Gaber, R. P. / Transport processes during sterilization of vertical and 5 degree horizontal dead-legs. In: Bioprocess Engineering. 1995 ; Vol. 12, No. 6. pp. 293-304.
@article{42ae0109fdb540769c6d9efd281cc1dd,
title = "Transport processes during sterilization of vertical and 5 degree horizontal dead-legs",
abstract = "Experimental temperature profiles and biological kill data from dead-ended tubes of various diameters and lengths commonly used in biotechnology applications were compared to those predicted by a finite element model for steam-in-place (SIP) sterilization at 122°C. Diameter was shown experimentally and numerically to have a significant effect with larger diameter tubes exhibiting greater buoyant driven convective flow and more rapid sterilization. The overall Grashof number was shown to be the significant parameter relating magnitude of convective flow to tube diameter and varied as the diameter cubed. Analysis of air/steam mixture flow patterns showed air displacement from 0.4 cm diameter tubes to be due primarily to molecular diffusion, whereas 1.0 and 1.7 cm tubes showed a two-stage convective flow pattern. There exists a critical diameter of 0.4 cm below which SIP sterilization due to buoyancy driven flow does not occur and steam bleeders should be used.",
author = "Young, {J. H.} and William Lasher and Gaber, {R. P.}",
year = "1995",
month = "5",
day = "1",
doi = "10.1007/BF00369506",
language = "English (US)",
volume = "12",
pages = "293--304",
journal = "Bioprocess and Biosystems Engineering",
issn = "1615-7591",
publisher = "Springer Verlag",
number = "6",

}

Transport processes during sterilization of vertical and 5 degree horizontal dead-legs. / Young, J. H.; Lasher, William; Gaber, R. P.

In: Bioprocess Engineering, Vol. 12, No. 6, 01.05.1995, p. 293-304.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Transport processes during sterilization of vertical and 5 degree horizontal dead-legs

AU - Young, J. H.

AU - Lasher, William

AU - Gaber, R. P.

PY - 1995/5/1

Y1 - 1995/5/1

N2 - Experimental temperature profiles and biological kill data from dead-ended tubes of various diameters and lengths commonly used in biotechnology applications were compared to those predicted by a finite element model for steam-in-place (SIP) sterilization at 122°C. Diameter was shown experimentally and numerically to have a significant effect with larger diameter tubes exhibiting greater buoyant driven convective flow and more rapid sterilization. The overall Grashof number was shown to be the significant parameter relating magnitude of convective flow to tube diameter and varied as the diameter cubed. Analysis of air/steam mixture flow patterns showed air displacement from 0.4 cm diameter tubes to be due primarily to molecular diffusion, whereas 1.0 and 1.7 cm tubes showed a two-stage convective flow pattern. There exists a critical diameter of 0.4 cm below which SIP sterilization due to buoyancy driven flow does not occur and steam bleeders should be used.

AB - Experimental temperature profiles and biological kill data from dead-ended tubes of various diameters and lengths commonly used in biotechnology applications were compared to those predicted by a finite element model for steam-in-place (SIP) sterilization at 122°C. Diameter was shown experimentally and numerically to have a significant effect with larger diameter tubes exhibiting greater buoyant driven convective flow and more rapid sterilization. The overall Grashof number was shown to be the significant parameter relating magnitude of convective flow to tube diameter and varied as the diameter cubed. Analysis of air/steam mixture flow patterns showed air displacement from 0.4 cm diameter tubes to be due primarily to molecular diffusion, whereas 1.0 and 1.7 cm tubes showed a two-stage convective flow pattern. There exists a critical diameter of 0.4 cm below which SIP sterilization due to buoyancy driven flow does not occur and steam bleeders should be used.

UR - http://www.scopus.com/inward/record.url?scp=0029025653&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0029025653&partnerID=8YFLogxK

U2 - 10.1007/BF00369506

DO - 10.1007/BF00369506

M3 - Article

VL - 12

SP - 293

EP - 304

JO - Bioprocess and Biosystems Engineering

JF - Bioprocess and Biosystems Engineering

SN - 1615-7591

IS - 6

ER -